SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\Routine\Scout AX RT or LT TA:7.3 s PAT:Off Voxel size:0.8×0.8×8.0 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 3 Dist. factor 50 % Position Isocenter Orientation Transversal Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % FoV read 400 mm FoV phase 100.0 % Slice thickness 8.0 mm TR 15.0 ms TE 5.00 ms Averages 1 Concatenations 3 Filter Distortion Corr.(2D), Elliptical filter Coil elements HW1-3 Contrast TD 0 ms MTC Off Magn. preparation None Flip angle 40 deg Fat suppr. None Water suppr. None SWI Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 50 % Phase partial Fourier Off Interpolation On PAT mode None Image Filter Off Distortion Corr. On TD 0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 3 Dist. factor 50 % Position Isocenter Phase enc. dir. A >> P Phase oversampling 0 % Multi-slice mode Sequential Series Interleaved Saturation mode Standard Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Tim CT mode Off System Body Off HW1 On HW2 On HW3 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SRFExcit 1H 185.368 V Gain High Table position 0 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Segments 1 Tagging None Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Phase stabilisation Off Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Off Contrasts 1 Bandwidth 180 Hz/Px Flow comp. No Allowed delay 0 s RF pulse type Fast Gradient mode Normal Excitation Slice-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements HW1-3 Acquisition duration 0 ms Mode Off BOLD Subtract Off Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\Routine\Scout Sag+Coro RT or LT TA:0:13 PAT:Off Voxel size:0.5×0.5×8.0 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 2 Slices 3 Dist. factor 50 % Position L12.4 A12.0 H10.0 mm Orientation S > T-35.6 > C1.5 Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % FoV read 250 mm FoV phase 100.0 % Slice thickness 8.0 mm TR 15.0 ms TE 5.00 ms Averages 1 Concatenations 6 Filter Distortion Corr.(2D), Elliptical filter Coil elements HW1-3 Contrast TD 0 ms MTC Off Magn. preparation None Flip angle 40 deg Fat suppr. None Water suppr. None SWI Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 50 % Phase partial Fourier Off Interpolation On PAT mode None Image Filter Off Distortion Corr. On TD 0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 2 Slices 3 Dist. factor 50 % Position L12.4 A12.0 H0.0 mm Phase enc. dir. A >> P Phase oversampling 0 % Multi-slice mode Sequential Series Interleaved Saturation mode Standard Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Tim CT mode Off System Body Off HW1 On HW2 On HW3 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Sum of Squares AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SRFExcit 1H 185.368 V Gain High Table position 10 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Segments 1 Tagging None Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Phase stabilisation Off Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Off Contrasts 1 Bandwidth 180 Hz/Px Flow comp. No Allowed delay 0 s RF pulse type Fast Gradient mode Normal Excitation Slice-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements HW1-3 Acquisition duration 0 ms Mode Off BOLD Subtract Off Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\Routine\T2 Axial SPAIR fs RT or LT TA:3:12 PAT:2 Voxel size:0.4×0.4×2.0 mm Rel. SNR:1.18 :tse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 20 Dist. factor 10 % Position R32.4 P1.9 H0.4 mm Orientation T > S20.1 > C-3.3 Phase enc. dir. R >> L AutoAlign --- Phase oversampling 0 % FoV read 100 mm FoV phase 100.0 % Slice thickness 2.0 mm TR 5000.0 ms TE 59.0 ms Averages 4 Concatenations 1 Filter Distortion Corr.(2D), Elliptical filter, B1 filter Coil elements HW1-3 Contrast MTC Off Magn. preparation None Flip angle 170 deg Fat suppr. SPAIR Fat sat. mode Weak Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 100 % Phase partial Fourier Allowed Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 32 Reference scan mode Self-calibration Image Filter Off Distortion Corr. On Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter On Unfiltered images Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 20 Dist. factor 10 % Position R32.4 P1.9 H4.4 mm Phase enc. dir. R >> L Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. SPAIR Water suppr. None Special sat. Parallel F/H Gap 10 mm Thickness 50 mm Fat sat. mode Weak Special sat. Parallel F/H Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off HW1 On HW2 On HW3 On Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R32.4 P1.9 H4.4 mm Rotation 90.00 deg A >> P 100 mm R >> L 100 mm F >> H 44 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 4 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 230 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 11.9 ms Define Turbo factor Turbo factor 15 Echo trains per slice 9 RF pulse type Normal Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements HW1-3 Acquisition duration 0 ms Mode Off BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\Routine\T1 Ax no fs RT or LT TA:3:36 PAT:Off Voxel size:0.3×0.3×2.0 mm Rel. SNR:1.00 :tse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 28 Dist. factor 10 % Position R30.5 P3.0 H0.4 mm Orientation Transversal Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % FoV read 100 mm FoV phase 62.5 % Slice thickness 2.0 mm TR 662.0 ms TE 26.0 ms Averages 2 Concatenations 4 Filter Distortion Corr.(2D), Elliptical filter, B1 filter Coil elements HW1-3 Contrast TD 0.0 ms MTC Off Magn. preparation None Flip angle 150 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 320 Phase resolution 100 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode None Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter On Unfiltered images Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 28 Dist. factor 10 % Position R30.5 P3.0 H4.4 mm Orientation T > S20.0 > C-2.9 Phase enc. dir. A >> P Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 1 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off HW1 On HW2 On HW3 On Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SLoop1RSatSS 1H 408.000 V Gain High Table position 4 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Short term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 260 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 13.2 ms Define Turbo factor Turbo factor 5 Echo trains per slice 40 RF pulse type Low SAR Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements HW1-3 Acquisition duration 0 ms Mode Off BOLD Subtract Off Save images Off Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\Routine\T2 spc Coro SPAIR iso TA:7:09 PAT:2 Voxel size:1.0×1.0×1.0 mm Rel. SNR:1.08 :spcR --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Position L7.2 P13.7 H0.3 mm Orientation Coronal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 100 % Slice oversampling 6.7 % FoV read 125 mm FoV phase 100.0 % Slice thickness 1.00 mm TR 1000 ms TE 99.0 ms Concatenations 1 Filter Raw filter, Distortion Corr.(2D), Prescan Normalize Coil elements NE1,2;SP1 Contrast MTC Off Magn. preparation None Flip angle 130 deg Fat suppr. SPAIR Fat sat. mode Strong Water suppr. None Restore magn. On Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 128 Phase resolution 100 % Phase partial Fourier Allowed Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Accel. factor 3D 1 Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Slice resolution 80 % Slice partial Fourier Off Geometry Nr. of slab groups 1 Slabs 1 Position L7.2 P13.7 F39.7 mm Phase enc. dir. R >> L Phase oversampling 100 % Slice oversampling 6.7 % Slices per slab 120 Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. SPAIR Water suppr. None Special sat. None Fat sat. mode Strong Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. On System Body Off HE1 Off HE3 Off NE1 On HE2 Off HE4 Off NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position L7.2 P13.7 F39.7 mm Rotation 0.00 deg F >> H 125 mm R >> L 125 mm A >> P 120 mm Frequency 1H 123.200000 MHz Correction factor 1 optfsOptfs 1H 135.123 V ! Gain High Table position 40 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Trigger delay 0 ms Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning Off Reordering Linear Bandwidth 501 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 4.31 ms Adiabatic-mode Off Turbo factor 47 Echo train duration 190 RF pulse type Normal Gradient mode Fast Excitation Slab-sel. Flip angle mode Constant TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements NE1,2;SP1 Acquisition duration 0 ms Organ under exam. None BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\Routine\PD COR fs LT or RT TA:6:37 PAT:Off Voxel size:0.3×0.3×2.0 mm Rel. SNR:1.00 :tse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 24 Dist. factor 10 % Position L86.3 A47.0 H0.3 mm Orientation C > S3.8 > T2.0 Phase enc. dir. R >> L AutoAlign --- Phase oversampling 0 % FoV read 100 mm FoV phase 100.0 % Slice thickness 2.0 mm TR 3240.0 ms TE 41.0 ms Averages 3 Concatenations 1 Filter Distortion Corr.(2D), Elliptical filter, B1 filter Coil elements HW1-3 Contrast MTC Off Magn. preparation None Flip angle 150 deg Fat suppr. Fat sat. Fat sat. mode Weak Water suppr. None Restore magn. Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Off Resolution Base resolution 320 Phase resolution 100 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode None Image Filter Off Distortion Corr. On Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter On Unfiltered images Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 24 Dist. factor 10 % Position L86.3 A47.0 H16.3 mm Phase enc. dir. R >> L Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. Fat sat. Water suppr. None Special sat. None Fat sat. mode Weak Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off HW1 On HW2 On HW3 On Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position L86.3 A47.0 H16.3 mm Rotation 31.40 deg F >> H 100 mm R >> L 100 mm A >> P 53 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 16 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Short term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 198 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 13.6 ms Define Turbo factor Turbo factor 8 Echo trains per slice 40 RF pulse type Normal Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements HW1-3 Acquisition duration 0 ms Mode Off BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\Routine\ep2d_dti_12 direction_p2 TA:6:16 PAT:2 Voxel size:2.7×2.7×4.0 mm Rel. SNR:1.00 :epse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 25 Dist. factor 0 % Position R81.1 P56.7 F0.1 mm Orientation Transversal Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % FoV read 260 mm FoV phase 81.6 % Slice thickness 4.0 mm TR 4800 ms TE 75.0 ms Averages 3 Concatenations 1 Filter Raw filter Coil elements BO1-3;SP6 Contrast MTC Off Magn. preparation None Fat suppr. Fat sat. Fat sat. mode Strong Averaging mode Long term Delay in TR 0 ms Reconstruction Magnitude Multiple series Off Resolution Base resolution 98 Phase resolution 100 % Phase partial Fourier 6/8 Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Separate Distortion Corr. Off Prescan Normalize Off Normalize Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Dynamic Field Corr. Off Geometry Nr. of slice groups 1 Slices 25 Dist. factor 0 % Position R81.1 P56.7 H83.9 mm Phase enc. dir. A >> P Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. Fat sat. Special sat. None Fat sat. mode Strong Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off BO1 On BO2 On BO3 On SP5 Off SP6 On SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Coil Combine Mode Sum of Squares AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V ! Position R97.5 P56.7 H83.9 mm ! Rotation 0.00 deg ! R >> L 144 mm ! A >> P 151 mm ! F >> H 119 mm Frequency 1H 123.200000 MHz Correction factor 1 AddCSaCSatNS 1H 83.814 V Gain High Table position 84 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Averaging mode Long term Multi-slice mode Interleaved Bandwidth 2126 Hz/Px Optimization None Free echo spacing Off Echo spacing 0.56 ms EPI factor 80 RF pulse type Low SAR Gradient mode Fast TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1-3;SP6 Acquisition duration 0 ms BOLD Delay in TR 0 ms Diffusion mode MDDW Diff. weightings 3 b-value 1 0 s/mm² Diff. weighted images On Trace weighted images On ADC maps On FA maps On Mosaic On Tensor On Distortion Corr. Off b-Value >= 0 s/mm² Exponential ADC Maps Off Invert Gray Scale Off Calculated Image Off Calculated bValue 1400 s/mm² SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\Routine\psif_fs_80b TA:3:58 PAT:Off Voxel size:0.9×0.9×1.0 mm Rel. SNR:1.00 :ps_rr --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position L51.1 P17.3 H0.5 mm Orientation S > T-30.1 > C11.9 Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % Slice oversampling 33.3 % FoV read 120 mm FoV phase 100.0 % Slice thickness 1.00 mm TR 12.00 ms TE 3.07 ms Averages 1 Filter Distortion Corr.(2D) Coil elements FS;SP1-3 Contrast Flip angle 35 deg Fat suppr. Water excit. normal Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 128 Phase resolution 100 % Phase partial Fourier Off Interpolation Off PAT mode None Image Filter Off Distortion Corr. On Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Slice resolution 80 % Slice partial Fourier Off Geometry Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position L51.1 P17.3 F24.5 mm Phase enc. dir. A >> P Phase oversampling 0 % Slice oversampling 33.3 % Slices per slab 144 Multi-slice mode Sequential Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. Water excit. normal Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off FS On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 On SP3 On SP4 Off Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position L51.1 P17.3 F24.5 mm Rotation 2.60 deg F >> H 120 mm A >> P 120 mm R >> L 144 mm Frequency 1H 123.200000 MHz Correction factor 1 ExcitWEs 0 1H 408.000 V Gain Low Table position 25 mm Img. Scale. Cor. 1.000 Physio Segments 1 Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning Off Averaging mode Long term Multi-slice mode Sequential Asymmetric echo Off Bandwidth 501 Hz/Px Flow comp. Read Optimization None RF pulse type Fast Gradient mode Fast* Excitation Slab-sel. TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements FS;SP1-3 Acquisition duration 0 ms Mode Off BOLD Diffusion mode Read Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\With Contrast\Scout AX RT or LT TA:7.3 s PAT:Off Voxel size:0.8×0.8×8.0 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 3 Dist. factor 50 % Position Isocenter Orientation Transversal Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % FoV read 400 mm FoV phase 100.0 % Slice thickness 8.0 mm TR 15.0 ms TE 5.00 ms Averages 1 Concatenations 3 Filter Distortion Corr.(2D), Elliptical filter Coil elements HW1-3 Contrast TD 0 ms MTC Off Magn. preparation None Flip angle 40 deg Fat suppr. None Water suppr. None SWI Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 50 % Phase partial Fourier Off Interpolation On PAT mode None Image Filter Off Distortion Corr. On TD 0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 3 Dist. factor 50 % Position Isocenter Phase enc. dir. A >> P Phase oversampling 0 % Multi-slice mode Sequential Series Interleaved Saturation mode Standard Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Tim CT mode Off System Body Off HW1 On HW2 On HW3 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SRFExcit 1H 185.368 V Gain High Table position 0 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Segments 1 Tagging None Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Phase stabilisation Off Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Off Contrasts 1 Bandwidth 180 Hz/Px Flow comp. No Allowed delay 0 s RF pulse type Fast Gradient mode Normal Excitation Slice-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements HW1-3 Acquisition duration 0 ms Mode Off BOLD Subtract Off Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\With Contrast\Scout Sag+Coro RT or LT TA:0:13 PAT:Off Voxel size:0.5×0.5×8.0 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 2 Slices 3 Dist. factor 50 % Position L12.4 A12.0 H10.0 mm Orientation S > T-35.6 > C1.5 Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % FoV read 250 mm FoV phase 100.0 % Slice thickness 8.0 mm TR 15.0 ms TE 5.00 ms Averages 1 Concatenations 6 Filter Distortion Corr.(2D), Elliptical filter Coil elements HW1-3 Contrast TD 0 ms MTC Off Magn. preparation None Flip angle 40 deg Fat suppr. None Water suppr. None SWI Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 50 % Phase partial Fourier Off Interpolation On PAT mode None Image Filter Off Distortion Corr. On TD 0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 2 Slices 3 Dist. factor 50 % Position L12.4 A12.0 H0.0 mm Phase enc. dir. A >> P Phase oversampling 0 % Multi-slice mode Sequential Series Interleaved Saturation mode Standard Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Tim CT mode Off System Body Off HW1 On HW2 On HW3 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Sum of Squares AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SRFExcit 1H 185.368 V Gain High Table position 10 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Segments 1 Tagging None Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Phase stabilisation Off Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Off Contrasts 1 Bandwidth 180 Hz/Px Flow comp. No Allowed delay 0 s RF pulse type Fast Gradient mode Normal Excitation Slice-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements HW1-3 Acquisition duration 0 ms Mode Off BOLD Subtract Off Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\With Contrast\T2 Axial SPAIR fs RT o r LT TA:3:12 PAT:2 Voxel size:0.4×0.4×2.0 mm Rel. SNR:1.18 :tse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 20 Dist. factor 10 % Position R32.4 P1.9 H0.4 mm Orientation T > S20.1 > C-3.3 Phase enc. dir. R >> L AutoAlign --- Phase oversampling 0 % FoV read 100 mm FoV phase 100.0 % Slice thickness 2.0 mm TR 5000.0 ms TE 59.0 ms Averages 4 Concatenations 1 Filter Distortion Corr.(2D), Elliptical filter, B1 filter Coil elements HW1-3 Contrast MTC Off Magn. preparation None Flip angle 170 deg Fat suppr. SPAIR Fat sat. mode Weak Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 100 % Phase partial Fourier Allowed Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 32 Reference scan mode Self-calibration Image Filter Off Distortion Corr. On Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter On Unfiltered images Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 20 Dist. factor 10 % Position R32.4 P1.9 H4.4 mm Phase enc. dir. R >> L Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. SPAIR Water suppr. None Special sat. Parallel F/H Gap 10 mm Thickness 50 mm Fat sat. mode Weak Special sat. Parallel F/H Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off HW1 On HW2 On HW3 On Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R32.4 P1.9 H4.4 mm Rotation 90.00 deg A >> P 100 mm R >> L 100 mm F >> H 44 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 4 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 230 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 11.9 ms Define Turbo factor Turbo factor 15 Echo trains per slice 9 RF pulse type Normal Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements HW1-3 Acquisition duration 0 ms Mode Off BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\With Contrast\T1 Ax no fs RT or LT TA:3:36 PAT:Off Voxel size:0.3×0.3×2.0 mm Rel. SNR:1.00 :tse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 28 Dist. factor 10 % Position R30.5 P3.0 H0.4 mm Orientation Transversal Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % FoV read 100 mm FoV phase 62.5 % Slice thickness 2.0 mm TR 662.0 ms TE 26.0 ms Averages 2 Concatenations 4 Filter Distortion Corr.(2D), Elliptical filter, B1 filter Coil elements HW1-3 Contrast TD 0.0 ms MTC Off Magn. preparation None Flip angle 150 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 320 Phase resolution 100 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode None Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter On Unfiltered images Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 28 Dist. factor 10 % Position R30.5 P3.0 H4.4 mm Orientation T > S20.0 > C-2.9 Phase enc. dir. A >> P Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 1 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off HW1 On HW2 On HW3 On Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SLoop1RSatSS 1H 408.000 V Gain High Table position 4 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Short term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 260 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 13.2 ms Define Turbo factor Turbo factor 5 Echo trains per slice 40 RF pulse type Low SAR Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements HW1-3 Acquisition duration 0 ms Mode Off BOLD Subtract Off Save images Off Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\With Contrast\T2 spc Coro SPAIR iso TA:7:09 PAT:2 Voxel size:1.0×1.0×1.0 mm Rel. SNR:1.08 :spcR --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Position L7.2 P13.7 H0.3 mm Orientation Coronal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 100 % Slice oversampling 6.7 % FoV read 125 mm FoV phase 100.0 % Slice thickness 1.00 mm TR 1000 ms TE 99.0 ms Concatenations 1 Filter Raw filter, Distortion Corr.(2D), Prescan Normalize Coil elements NE1,2;SP1 Contrast MTC Off Magn. preparation None Flip angle 130 deg Fat suppr. SPAIR Fat sat. mode Strong Water suppr. None Restore magn. On Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 128 Phase resolution 100 % Phase partial Fourier Allowed Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Accel. factor 3D 1 Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Slice resolution 80 % Slice partial Fourier Off Geometry Nr. of slab groups 1 Slabs 1 Position L7.2 P13.7 F39.7 mm Phase enc. dir. R >> L Phase oversampling 100 % Slice oversampling 6.7 % Slices per slab 120 Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. SPAIR Water suppr. None Special sat. None Fat sat. mode Strong Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. On System Body Off HE1 Off HE3 Off NE1 On HE2 Off HE4 Off NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position L7.2 P13.7 F39.7 mm Rotation 0.00 deg F >> H 125 mm R >> L 125 mm A >> P 120 mm Frequency 1H 123.200000 MHz Correction factor 1 optfsOptfs 1H 135.123 V ! Gain High Table position 40 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Trigger delay 0 ms Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning Off Reordering Linear Bandwidth 501 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 4.31 ms Adiabatic-mode Off Turbo factor 47 Echo train duration 190 RF pulse type Normal Gradient mode Fast Excitation Slab-sel. Flip angle mode Constant TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements NE1,2;SP1 Acquisition duration 0 ms Organ under exam. None BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\With Contrast\PD COR fs LT or RT TA:6:37 PAT:Off Voxel size:0.3×0.3×2.0 mm Rel. SNR:1.00 :tse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 24 Dist. factor 10 % Position L86.3 A47.0 H0.3 mm Orientation C > S3.8 > T2.0 Phase enc. dir. R >> L AutoAlign --- Phase oversampling 0 % FoV read 100 mm FoV phase 100.0 % Slice thickness 2.0 mm TR 3240.0 ms TE 41.0 ms Averages 3 Concatenations 1 Filter Distortion Corr.(2D), Elliptical filter, B1 filter Coil elements HW1-3 Contrast MTC Off Magn. preparation None Flip angle 150 deg Fat suppr. Fat sat. Fat sat. mode Weak Water suppr. None Restore magn. Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Off Resolution Base resolution 320 Phase resolution 100 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode None Image Filter Off Distortion Corr. On Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter On Unfiltered images Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 24 Dist. factor 10 % Position L86.3 A47.0 H16.3 mm Phase enc. dir. R >> L Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. Fat sat. Water suppr. None Special sat. None Fat sat. mode Weak Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off HW1 On HW2 On HW3 On Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position L86.3 A47.0 H16.3 mm Rotation 31.40 deg F >> H 100 mm R >> L 100 mm A >> P 53 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 16 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Short term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 198 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 13.6 ms Define Turbo factor Turbo factor 8 Echo trains per slice 40 RF pulse type Normal Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements HW1-3 Acquisition duration 0 ms Mode Off BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\With Contrast\ep2d_dti_12 direction_p2 TA:6:16 PAT:2 Voxel size:2.7×2.7×4.0 mm Rel. SNR:1.00 :epse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 25 Dist. factor 0 % Position R81.1 P56.7 F0.1 mm Orientation Transversal Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % FoV read 260 mm FoV phase 81.6 % Slice thickness 4.0 mm TR 4800 ms TE 75.0 ms Averages 3 Concatenations 1 Filter Raw filter Coil elements BO1-3;SP6 Contrast MTC Off Magn. preparation None Fat suppr. Fat sat. Fat sat. mode Strong Averaging mode Long term Delay in TR 0 ms Reconstruction Magnitude Multiple series Off Resolution Base resolution 98 Phase resolution 100 % Phase partial Fourier 6/8 Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Separate Distortion Corr. Off Prescan Normalize Off Normalize Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Dynamic Field Corr. Off Geometry Nr. of slice groups 1 Slices 25 Dist. factor 0 % Position R81.1 P56.7 H83.9 mm Phase enc. dir. A >> P Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. Fat sat. Special sat. None Fat sat. mode Strong Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off BO1 On BO2 On BO3 On SP5 Off SP6 On SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Coil Combine Mode Sum of Squares AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V ! Position R97.5 P56.7 H83.9 mm ! Rotation 0.00 deg ! R >> L 144 mm ! A >> P 151 mm ! F >> H 119 mm Frequency 1H 123.200000 MHz Correction factor 1 AddCSaCSatNS 1H 83.814 V Gain High Table position 84 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Averaging mode Long term Multi-slice mode Interleaved Bandwidth 2126 Hz/Px Optimization None Free echo spacing Off Echo spacing 0.56 ms EPI factor 80 RF pulse type Low SAR Gradient mode Fast TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1-3;SP6 Acquisition duration 0 ms BOLD Delay in TR 0 ms Diffusion mode MDDW Diff. weightings 3 b-value 1 0 s/mm² Diff. weighted images On Trace weighted images On ADC maps On FA maps On Mosaic On Tensor On Distortion Corr. Off b-Value >= 0 s/mm² Exponential ADC Maps Off Invert Gray Scale Off Calculated Image Off Calculated bValue 1400 s/mm² SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\With Contrast\psif_fs_80b TA:3:58 PAT:Off Voxel size:0.9×0.9×1.0 mm Rel. SNR:1.00 :ps_rr --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position L51.1 P17.3 H0.5 mm Orientation S > T-30.1 > C11.9 Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % Slice oversampling 33.3 % FoV read 120 mm FoV phase 100.0 % Slice thickness 1.00 mm TR 12.00 ms TE 3.07 ms Averages 1 Filter Distortion Corr.(2D) Coil elements FS;SP1-3 Contrast Flip angle 35 deg Fat suppr. Water excit. normal Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 128 Phase resolution 100 % Phase partial Fourier Off Interpolation Off PAT mode None Image Filter Off Distortion Corr. On Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Slice resolution 80 % Slice partial Fourier Off Geometry Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position L51.1 P17.3 F24.5 mm Phase enc. dir. A >> P Phase oversampling 0 % Slice oversampling 33.3 % Slices per slab 144 Multi-slice mode Sequential Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. Water excit. normal Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off FS On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 On SP3 On SP4 Off Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position L51.1 P17.3 F24.5 mm Rotation 2.60 deg F >> H 120 mm A >> P 120 mm R >> L 144 mm Frequency 1H 123.200000 MHz Correction factor 1 ExcitWEs 0 1H 408.000 V Gain Low Table position 25 mm Img. Scale. Cor. 1.000 Physio Segments 1 Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning Off Averaging mode Long term Multi-slice mode Sequential Asymmetric echo Off Bandwidth 501 Hz/Px Flow comp. Read Optimization None RF pulse type Fast Gradient mode Fast* Excitation Slab-sel. TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements FS;SP1-3 Acquisition duration 0 ms Mode Off BOLD Diffusion mode Read Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\With Contrast\T1vibe Ax Coro TA:2:55 PAT:2 Voxel size:0.6×0.6×0.6 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position R7.9 P43.8 H0.0 mm Orientation Coronal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 23 % Slice oversampling 44.4 % FoV read 100 mm FoV phase 75.0 % Slice thickness 0.60 mm TR 12.70 ms TE 4.49 ms Averages 1 Concatenations 1 Filter Distortion Corr.(2D), Prescan Normalize Coil elements HW1-3 Contrast Flip angle 10.0 deg Fat suppr. Water excit. normal Water suppr. None Dixon No Dixon Save original images On Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Off Resolution Base resolution 160 Phase resolution 100 % Phase partial Fourier Off Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Accel. factor 3D 1 Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter Off POCS Off Slice resolution 100 % Slice partial Fourier 7/8 Geometry Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position R7.9 P43.8 H13.0 mm Phase enc. dir. R >> L Phase oversampling 23 % Slice oversampling 44.4 % Slices per slab 144 Multi-slice mode Sequential Series Ascending Nr. of sat. regions 0 Position mode L-P-H Fat suppr. Water excit. normal Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off HW1 On HW2 On HW3 On Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R7.9 P43.8 H13.0 mm Rotation 0.00 deg F >> H 100 mm R >> L 75 mm A >> P 87 mm Frequency 1H 123.200000 MHz Correction factor 1 ExcitWEns 0 1H 15.741 V Gain Low Table position 13 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning On Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Weak Contrasts 1 Bandwidth 150 Hz/Px Optimization Min. TE Allowed delay 0 s RF pulse type Normal Gradient mode Normal Excitation Non-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements HW1-3 Acquisition duration 0 ms Mode Off BOLD 3D centric reordering Off Time to center 83.0 s Subtract Off Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Carpal Tunnel\With Contrast\T1vibe Coro Post TA:2:55 PAT:2 Voxel size:0.6×0.6×0.6 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position R7.9 P43.8 H0.0 mm Orientation Coronal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 23 % Slice oversampling 44.4 % FoV read 100 mm FoV phase 75.0 % Slice thickness 0.60 mm TR 12.70 ms TE 4.49 ms Averages 1 Concatenations 1 Filter Distortion Corr.(2D), Prescan Normalize Coil elements HW1-3 Contrast Flip angle 10.0 deg Fat suppr. Water excit. normal Water suppr. None Dixon No Dixon Save original images On Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Off Resolution Base resolution 160 Phase resolution 100 % Phase partial Fourier Off Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Accel. factor 3D 1 Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter Off POCS Off Slice resolution 100 % Slice partial Fourier 7/8 Geometry Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position R7.9 P43.8 H13.0 mm Phase enc. dir. R >> L Phase oversampling 23 % Slice oversampling 44.4 % Slices per slab 144 Multi-slice mode Sequential Series Ascending Nr. of sat. regions 0 Position mode L-P-H Fat suppr. Water excit. normal Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off HW1 On HW2 On HW3 On Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R7.9 P43.8 H13.0 mm Rotation 0.00 deg F >> H 100 mm R >> L 75 mm A >> P 87 mm Frequency 1H 123.200000 MHz Correction factor 1 ExcitWEns 0 1H 15.741 V Gain Low Table position 13 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning On Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Weak Contrasts 1 Bandwidth 150 Hz/Px Optimization Min. TE Allowed delay 0 s RF pulse type Normal Gradient mode Normal Excitation Non-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements HW1-3 Acquisition duration 0 ms Mode Off BOLD 3D centric reordering Off Time to center 83.0 s Subtract Off Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\Routine\localizer TA:0:23 PAT:2 Voxel size:2.0×2.0×8.0 mm Rel. SNR:1.00 :h --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display On Wait for user to start Off Start measurements single Routine Nr. of slice groups 3 Slices 10 Dist. factor 150 % Position L0.0 P0.0 F50.0 mm Orientation Transversal Phase enc. dir. A >> P AutoAlign --- Phase oversampling 30 % FoV read 500 mm FoV phase 100.0 % Slice thickness 8.0 mm TR 1000.0 ms TE 82.0 ms Averages 1 Concatenations 1 Filter Distortion Corr.(2D), Elliptical filter Coil elements BO1-3;HE1-4;NE1,2;SP1,2 Contrast MTC Off Magn. preparation None Flip angle 180 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 70 % Phase partial Fourier 5/8 Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 3 Slices 10 Dist. factor 150 % Position L0.0 P0.0 F50.0 mm Phase enc. dir. A >> P Phase oversampling 30 % Multi-slice mode Single shot Series Descending Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 On BO3 On HE1 On HE3 On NE1 On HE2 On HE4 On NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 On SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 202.751 V Gain High Table position 0 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Resp. control Breath-hold Inline Inline Composing Off Distortion correction Off Sequence Introduction Off Dimension 2D Averaging mode Long term Multi-slice mode Single shot Contrasts 1 Bandwidth 476 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 5.1 ms Turbo factor 179 RF pulse type Normal Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1-3;HE1-4;NE1,2;SP1,2 Acquisition duration 0 ms Mode Min flip angle Min flip angle 150 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\Routine\T2spc Sag C-Spine TA:5:11 PAT:2 Voxel size:0.5×0.5×0.6 mm Rel. SNR:1.00 :spcR --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Position R8.5 P14.5 F0.3 mm Orientation S > T2.5 Phase enc. dir. H >> F AutoAlign --- Phase oversampling 80 % Slice oversampling 12.5 % FoV read 250 mm FoV phase 100.0 % Slice thickness 0.60 mm TR 1500 ms TE 136.0 ms Concatenations 1 Filter Raw filter, Distortion Corr.(3D) Coil elements BO1;HE3,4;NE1,2;SP1 Contrast MTC Off Magn. preparation None Flip angle 140 deg Fat suppr. None Water suppr. None Restore magn. On Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 100 % Phase partial Fourier Allowed Interpolation On PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Accel. factor 3D 1 Mode 3D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Slice resolution 100 % Slice partial Fourier 6/8 Geometry Nr. of slab groups 1 Slabs 1 Position R8.5 P14.5 H9.7 mm Phase enc. dir. H >> F Phase oversampling 80 % Slice oversampling 12.5 % Slices per slab 64 Series Ascending Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. On System Body Off BO1 On BO2 Off BO3 Off HE1 Off HE3 On NE1 On HE2 Off HE4 On NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R8.5 P14.5 H9.7 mm Rotation 90.00 deg A >> P 250 mm F >> H 250 mm R >> L 39 mm Frequency 1H 123.200000 MHz Correction factor 1 TSE_DEFT 1H 283.333 V ! Gain High Table position 10 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Trigger delay 0 ms Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning Off Reordering Linear Bandwidth 630 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 5.75 ms Adiabatic-mode On Turbo factor 88 Echo train duration 357 RF pulse type Normal Gradient mode Fast Excitation Slab-sel. Flip angle mode Constant TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;HE3,4;NE1,2;SP1 Acquisition duration 0 ms Organ under exam. None BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 3D Unfiltered images Off Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\Routine\T1 Ax TA:2:35 PAT:2 Voxel size:0.6×0.6×3.0 mm Rel. SNR:1.00 :tse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 38 Dist. factor 20 % Position R6.1 P21.8 H0.2 mm Orientation Transversal Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % FoV read 330 mm FoV phase 81.3 % Slice thickness 3.0 mm TR 680.0 ms TE 9.1 ms Averages 2 Concatenations 2 Filter Distortion Corr.(2D) Coil elements BO1;NE1,2;SP1 Contrast TD 0.0 ms MTC Off Magn. preparation None Flip angle 180 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Off Resolution Base resolution 512 Phase resolution 80 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 32 Reference scan mode Self-calibration Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Geometry Nr. of slice groups 1 Slices 38 Dist. factor 20 % Position R6.1 P21.8 F21.8 mm Phase enc. dir. A >> P Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 Off BO3 Off HE1 Off HE3 Off NE1 On HE2 Off HE4 Off NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 22 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 296 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 9.1 ms Define Turbo factor Turbo factor 3 Echo trains per slice 56 RF pulse type Normal Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;NE1,2;SP1 Acquisition duration 0 ms Mode Min flip angle Min flip angle 150 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\Routine\STIRspc Coro TA:4:23 PAT:3 Voxel size:1.1×1.1×1.0 mm Rel. SNR:1.00 :spcir --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Position R10.9 P19.4 H0.2 mm Orientation C > T7.9 Phase enc. dir. R >> L AutoAlign --- Phase oversampling 70 % Slice oversampling 9.1 % FoV read 270 mm FoV phase 100.0 % Slice thickness 1.00 mm TR 1500 ms TE 160.0 ms Concatenations 1 Filter Raw filter, Distortion Corr.(3D) Coil elements BO1,2;HE3,4;NE1,2;SP1,2 Contrast MTC Off Magn. preparation Non-sel. IR TI 220 ms Fat suppr. None Water suppr. None Restore magn. Off Measurements 1 Reconstruction Magnitude Multiple series Each measurement Gradient 1st moment in read 200 mT ms^2/m Gradient 1st moment in phase 200 mT ms^2/m Gradient 1st moment in slice 0 mT ms^2/m Resolution Base resolution 256 Phase resolution 95 % Phase partial Fourier Allowed Interpolation Off PAT mode GRAPPA Accel. factor PE 3 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Accel. factor 3D 1 Mode 3D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Slice resolution 100 % Slice partial Fourier 6/8 Geometry Nr. of slab groups 1 Slabs 1 Position R10.9 P19.4 H1.2 mm Phase enc. dir. R >> L Phase oversampling 70 % Slice oversampling 9.1 % Slices per slab 88 Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off System Body Off BO1 On BO2 On BO3 Off HE1 Off HE3 On NE1 On HE2 Off HE4 On NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 On SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SLoopIRns1 1H 509.370 V ! Gain High Table position 1 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Trigger delay 0 ms Magn. preparation Non-sel. IR TI 220 ms Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning Off Reordering Linear Bandwidth 651 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 4.51 ms Adiabatic-mode On Turbo factor 128 Echo train duration 406 RF pulse type Fast Gradient mode Fast Excitation Slab-sel. Flip angle mode T2 var TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;HE3,4;NE1,2;SP1,2 Acquisition duration 0 ms Organ under exam. None BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 3D Unfiltered images Off Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\Routine\STIR Sag Left TA:3:47 PAT:2 Voxel size:0.7×0.7×3.0 mm Rel. SNR:1.00 :tir_rr --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 45 Dist. factor 10 % Position L59.3 P21.8 F0.1 mm Orientation S > T3.6 Phase enc. dir. A >> P AutoAlign --- Phase oversampling 40 % FoV read 220 mm FoV phase 100.0 % Slice thickness 3.0 mm TR 3000.0 ms TE 35.0 ms Averages 2 Concatenations 3 Filter Distortion Corr.(2D), Elliptical filter Coil elements BO1,2;HE3,4;NE1,2;SP1 Contrast TD 0.0 ms MTC Off Magn. preparation Slice-sel. IR TI 220 ms Freeze suppressed tissue Off Flip angle 160 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 320 Phase resolution 75 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 32 Reference scan mode Self-calibration Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 45 Dist. factor 10 % Position L59.3 P21.8 F6.1 mm Phase enc. dir. A >> P Phase oversampling 40 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 On BO3 Off HE1 Off HE3 On NE1 On HE2 Off HE4 On NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 6 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation Slice-sel. IR TI 220 ms Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 300 Hz/Px Flow comp. Read Allowed delay 60 s Echo spacing 8.82 ms Define Turbo factor Turbo factor 15 Echo trains per slice 12 RF pulse type Normal Gradient mode Normal Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;HE3,4;NE1,2;SP1 Acquisition duration 0 ms Mode Min flip angle Min flip angle 130 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\Routine\STIR Sag Right TA:3:47 PAT:2 Voxel size:0.7×0.7×3.0 mm Rel. SNR:1.00 :tir_rr --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 45 Dist. factor 10 % Position R78.0 P21.8 H0.3 mm Orientation S > T3.6 Phase enc. dir. A >> P AutoAlign --- Phase oversampling 40 % FoV read 220 mm FoV phase 100.0 % Slice thickness 3.0 mm TR 3000.0 ms TE 35.0 ms Averages 2 Concatenations 3 Filter Distortion Corr.(2D), Elliptical filter Coil elements BO1;HE3,4;NE1,2;SP1 Contrast TD 0.0 ms MTC Off Magn. preparation Slice-sel. IR TI 220 ms Freeze suppressed tissue Off Flip angle 160 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 320 Phase resolution 75 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 32 Reference scan mode Self-calibration Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 45 Dist. factor 10 % Position R78.0 P21.8 H3.3 mm Phase enc. dir. A >> P Phase oversampling 40 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 Off BO3 Off HE1 Off HE3 On NE1 On HE2 Off HE4 On NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 3 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation Slice-sel. IR TI 220 ms Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 300 Hz/Px Flow comp. Read Allowed delay 60 s Echo spacing 8.82 ms Define Turbo factor Turbo factor 15 Echo trains per slice 12 RF pulse type Normal Gradient mode Normal Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;HE3,4;NE1,2;SP1 Acquisition duration 0 ms Mode Min flip angle Min flip angle 130 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\Routine\DTI Ax TA:4:11 PAT:2 Voxel size:2.5×2.5×5.0 mm Rel. SNR:1.00 :epse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 30 Dist. factor 0 % Position R12.9 P29.6 F0.4 mm Orientation Transversal Phase enc. dir. P >> A AutoAlign --- Phase oversampling 0 % FoV read 320 mm FoV phase 81.5 % Slice thickness 5.0 mm TR 4700 ms TE 69.0 ms Averages 2 Concatenations 1 Filter Raw filter Coil elements BO2,3;SP4,5 Contrast MTC Off Magn. preparation None Fat suppr. SPAIR Fat sat. mode Strong Averaging mode Long term Delay in TR 0 ms Reconstruction Magnitude Multiple series Off Resolution Base resolution 130 Phase resolution 100 % Phase partial Fourier Off Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Separate Distortion Corr. Off Prescan Normalize Off Normalize Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Dynamic Field Corr. Off Geometry Nr. of slice groups 1 Slices 30 Dist. factor 0 % Position R12.9 P29.6 F130.4 mm Phase enc. dir. P >> A Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. SPAIR Special sat. None Fat sat. mode Strong Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off BO1 Off BO2 On BO3 On BO1 Off BO2 Off BO3 Off SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal R >> L Coronal A >> P Transversal F >> H Coil Combine Mode Sum of Squares AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R12.9 P29.6 F130.4 mm Rotation -180.00 deg R >> L 320 mm A >> P 261 mm F >> H 150 mm Frequency 1H 123.200000 MHz Correction factor 1 ExtExciteRF 1H 140.463 V Gain High Table position 130 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Averaging mode Long term Multi-slice mode Interleaved Bandwidth 2136 Hz/Px Optimization None Free echo spacing Off Echo spacing 0.59 ms EPI factor 106 RF pulse type Normal Gradient mode Fast* TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO2,3;SP4,5 Acquisition duration 0 ms BOLD Delay in TR 0 ms Diffusion mode MDDW Diff. weightings 3 b-value 1 0 s/mm² Diff. weighted images On Trace weighted images On ADC maps On FA maps On Mosaic On Tensor On Distortion Corr. Off b-Value >= 0 s/mm² Exponential ADC Maps Off Invert Gray Scale Off Calculated Image Off Calculated bValue 1400 s/mm² SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\Routine\T1 Coro VIBE Pre TA:3:37 PAT:4 Voxel size:0.9×0.9×1.0 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position R10.8 P15.7 H0.4 mm Orientation C > T7.9 Phase enc. dir. R >> L AutoAlign --- Phase oversampling 30 % Slice oversampling 66.7 % FoV read 270 mm FoV phase 100.0 % Slice thickness 1.00 mm TR 6.48 ms TE 3.05 ms Averages 5 Concatenations 1 Filter Distortion Corr.(3D), Elliptical filter Coil elements BO1,2;HE3,4;NE1,2;SP1,2 Contrast Flip angle 12.0 deg Fat suppr. Water excit. normal Water suppr. None Dixon No Dixon Save original images On Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 288 Phase resolution 100 % Phase partial Fourier Off Interpolation Off PAT mode GRAPPA Accel. factor PE 4 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Accel. factor 3D 1 Mode 3D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane POCS Off Slice resolution 50 % Slice partial Fourier Off Geometry Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position R10.8 P15.7 H2.4 mm Phase enc. dir. R >> L Phase oversampling 30 % Slice oversampling 66.7 % Slices per slab 72 Multi-slice mode Sequential Series Ascending Nr. of sat. regions 0 Position mode L-P-H Fat suppr. Water excit. normal Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off BO1 On BO2 On BO3 Off HE1 Off HE3 On NE1 On HE2 Off HE4 On NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 On SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R10.8 P15.7 H2.4 mm Rotation 0.00 deg F >> H 270 mm R >> L 270 mm A >> P 72 mm Frequency 1H 123.200000 MHz Correction factor 1 ExcitWEs 0 1H 70.015 V Gain Low Table position 2 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning Off Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Weak Contrasts 1 Bandwidth 500 Hz/Px Optimization Min. TE Allowed delay 30 s RF pulse type Fast Gradient mode Normal Excitation Slab-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;HE3,4;NE1,2;SP1,2 Acquisition duration 0 ms Mode Off BOLD 3D centric reordering Off Time to center 23.4 s Subtract On Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 3D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\Routine\T1 Coro VIBE Post TA:3:37 PAT:4 Voxel size:0.9×0.9×1.0 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position R10.8 P15.7 H0.4 mm Orientation C > T7.9 Phase enc. dir. R >> L AutoAlign --- Phase oversampling 30 % Slice oversampling 66.7 % FoV read 270 mm FoV phase 100.0 % Slice thickness 1.00 mm TR 6.48 ms TE 3.05 ms Averages 5 Concatenations 1 Filter Distortion Corr.(3D), Elliptical filter Coil elements BO1,2;HE3,4;NE1,2;SP1,2 Contrast Flip angle 12.0 deg Fat suppr. Water excit. normal Water suppr. None Dixon No Dixon Save original images On Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 288 Phase resolution 100 % Phase partial Fourier Off Interpolation Off PAT mode GRAPPA Accel. factor PE 4 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Accel. factor 3D 1 Mode 3D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane POCS Off Slice resolution 50 % Slice partial Fourier Off Geometry Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position R10.8 P15.7 H2.4 mm Phase enc. dir. R >> L Phase oversampling 30 % Slice oversampling 66.7 % Slices per slab 72 Multi-slice mode Sequential Series Ascending Nr. of sat. regions 0 Position mode L-P-H Fat suppr. Water excit. normal Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off BO1 On BO2 On BO3 Off HE1 Off HE3 On NE1 On HE2 Off HE4 On NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 On SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R10.8 P15.7 H2.4 mm Rotation 0.00 deg F >> H 270 mm R >> L 270 mm A >> P 72 mm Frequency 1H 123.200000 MHz Correction factor 1 ExcitWEs 0 1H 70.015 V Gain Low Table position 2 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning Off Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Weak Contrasts 1 Bandwidth 500 Hz/Px Optimization Min. TE Allowed delay 30 s RF pulse type Fast Gradient mode Normal Excitation Slab-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;HE3,4;NE1,2;SP1,2 Acquisition duration 0 ms Mode Off BOLD 3D centric reordering Off Time to center 23.4 s Subtract On Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 3D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\With Contrast\localizer TA:0:23 PAT:2 Voxel size:2.0×2.0×8.0 mm Rel. SNR:1.00 :h --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display On Wait for user to start Off Start measurements single Routine Nr. of slice groups 3 Slices 10 Dist. factor 150 % Position L0.0 P0.0 F50.0 mm Orientation Transversal Phase enc. dir. A >> P AutoAlign --- Phase oversampling 30 % FoV read 500 mm FoV phase 100.0 % Slice thickness 8.0 mm TR 1000.0 ms TE 82.0 ms Averages 1 Concatenations 1 Filter Distortion Corr.(2D), Elliptical filter Coil elements BO1-3;HE1-4;NE1,2;SP1,2 Contrast MTC Off Magn. preparation None Flip angle 180 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 70 % Phase partial Fourier 5/8 Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 3 Slices 10 Dist. factor 150 % Position L0.0 P0.0 F50.0 mm Phase enc. dir. A >> P Phase oversampling 30 % Multi-slice mode Single shot Series Descending Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 On BO3 On HE1 On HE3 On NE1 On HE2 On HE4 On NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 On SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 202.751 V Gain High Table position 0 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Resp. control Breath-hold Inline Inline Composing Off Distortion correction Off Sequence Introduction Off Dimension 2D Averaging mode Long term Multi-slice mode Single shot Contrasts 1 Bandwidth 476 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 5.1 ms Turbo factor 179 RF pulse type Normal Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1-3;HE1-4;NE1,2;SP1,2 Acquisition duration 0 ms Mode Min flip angle Min flip angle 150 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\With Contrast\T2spc Sag C-Spine TA:5:11 PAT:2 Voxel size:0.5×0.5×0.6 mm Rel. SNR:1.00 :spcR --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Position R8.5 P14.5 F0.3 mm Orientation S > T2.5 Phase enc. dir. H >> F AutoAlign --- Phase oversampling 80 % Slice oversampling 12.5 % FoV read 250 mm FoV phase 100.0 % Slice thickness 0.60 mm TR 1500 ms TE 136.0 ms Concatenations 1 Filter Raw filter, Distortion Corr.(3D) Coil elements BO1;HE3,4;NE1,2;SP1 Contrast MTC Off Magn. preparation None Flip angle 140 deg Fat suppr. None Water suppr. None Restore magn. On Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 100 % Phase partial Fourier Allowed Interpolation On PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Accel. factor 3D 1 Mode 3D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Slice resolution 100 % Slice partial Fourier 6/8 Geometry Nr. of slab groups 1 Slabs 1 Position R8.5 P14.5 H9.7 mm Phase enc. dir. H >> F Phase oversampling 80 % Slice oversampling 12.5 % Slices per slab 64 Series Ascending Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. On System Body Off BO1 On BO2 Off BO3 Off HE1 Off HE3 On NE1 On HE2 Off HE4 On NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R8.5 P14.5 H9.7 mm Rotation 90.00 deg A >> P 250 mm F >> H 250 mm R >> L 39 mm Frequency 1H 123.200000 MHz Correction factor 1 TSE_DEFT 1H 283.333 V ! Gain High Table position 10 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Trigger delay 0 ms Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning Off Reordering Linear Bandwidth 630 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 5.75 ms Adiabatic-mode On Turbo factor 88 Echo train duration 357 RF pulse type Normal Gradient mode Fast Excitation Slab-sel. Flip angle mode Constant TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;HE3,4;NE1,2;SP1 Acquisition duration 0 ms Organ under exam. None BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 3D Unfiltered images Off Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\With Contrast\T1 Ax TA:2:35 PAT:2 Voxel size:0.6×0.6×3.0 mm Rel. SNR:1.00 :tse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 38 Dist. factor 20 % Position R6.1 P21.8 H0.2 mm Orientation Transversal Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % FoV read 330 mm FoV phase 81.3 % Slice thickness 3.0 mm TR 680.0 ms TE 9.1 ms Averages 2 Concatenations 2 Filter Distortion Corr.(2D) Coil elements BO1;NE1,2;SP1 Contrast TD 0.0 ms MTC Off Magn. preparation None Flip angle 180 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Off Resolution Base resolution 512 Phase resolution 80 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 32 Reference scan mode Self-calibration Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Geometry Nr. of slice groups 1 Slices 38 Dist. factor 20 % Position R6.1 P21.8 F21.8 mm Phase enc. dir. A >> P Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 Off BO3 Off HE1 Off HE3 Off NE1 On HE2 Off HE4 Off NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 22 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 296 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 9.1 ms Define Turbo factor Turbo factor 3 Echo trains per slice 56 RF pulse type Normal Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;NE1,2;SP1 Acquisition duration 0 ms Mode Min flip angle Min flip angle 150 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\With Contrast\STIRspc Coro TA:4:23 PAT:3 Voxel size:1.1×1.1×1.0 mm Rel. SNR:1.00 :spcir --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Position R10.9 P19.4 H0.2 mm Orientation C > T7.9 Phase enc. dir. R >> L AutoAlign --- Phase oversampling 70 % Slice oversampling 9.1 % FoV read 270 mm FoV phase 100.0 % Slice thickness 1.00 mm TR 1500 ms TE 160.0 ms Concatenations 1 Filter Raw filter, Distortion Corr.(3D) Coil elements BO1,2;HE3,4;NE1,2;SP1,2 Contrast MTC Off Magn. preparation Non-sel. IR TI 220 ms Fat suppr. None Water suppr. None Restore magn. Off Measurements 1 Reconstruction Magnitude Multiple series Each measurement Gradient 1st moment in read 200 mT ms^2/m Gradient 1st moment in phase 200 mT ms^2/m Gradient 1st moment in slice 0 mT ms^2/m Resolution Base resolution 256 Phase resolution 95 % Phase partial Fourier Allowed Interpolation Off PAT mode GRAPPA Accel. factor PE 3 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Accel. factor 3D 1 Mode 3D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Slice resolution 100 % Slice partial Fourier 6/8 Geometry Nr. of slab groups 1 Slabs 1 Position R10.9 P19.4 H1.2 mm Phase enc. dir. R >> L Phase oversampling 70 % Slice oversampling 9.1 % Slices per slab 88 Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off System Body Off BO1 On BO2 On BO3 Off HE1 Off HE3 On NE1 On HE2 Off HE4 On NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 On SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SLoopIRns1 1H 509.370 V ! Gain High Table position 1 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Trigger delay 0 ms Magn. preparation Non-sel. IR TI 220 ms Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning Off Reordering Linear Bandwidth 651 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 4.51 ms Adiabatic-mode On Turbo factor 128 Echo train duration 406 RF pulse type Fast Gradient mode Fast Excitation Slab-sel. Flip angle mode T2 var TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;HE3,4;NE1,2;SP1,2 Acquisition duration 0 ms Organ under exam. None BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 3D Unfiltered images Off Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\With Contrast\STIR Sag Left TA:3:47 PAT:2 Voxel size:0.7×0.7×3.0 mm Rel. SNR:1.00 :tir_rr --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 45 Dist. factor 10 % Position L59.3 P21.8 F0.1 mm Orientation S > T3.6 Phase enc. dir. A >> P AutoAlign --- Phase oversampling 40 % FoV read 220 mm FoV phase 100.0 % Slice thickness 3.0 mm TR 3000.0 ms TE 35.0 ms Averages 2 Concatenations 3 Filter Distortion Corr.(2D), Elliptical filter Coil elements BO1,2;HE3,4;NE1,2;SP1 Contrast TD 0.0 ms MTC Off Magn. preparation Slice-sel. IR TI 220 ms Freeze suppressed tissue Off Flip angle 160 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 320 Phase resolution 75 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 32 Reference scan mode Self-calibration Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 45 Dist. factor 10 % Position L59.3 P21.8 F6.1 mm Phase enc. dir. A >> P Phase oversampling 40 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 On BO3 Off HE1 Off HE3 On NE1 On HE2 Off HE4 On NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 6 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation Slice-sel. IR TI 220 ms Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 300 Hz/Px Flow comp. Read Allowed delay 60 s Echo spacing 8.82 ms Define Turbo factor Turbo factor 15 Echo trains per slice 12 RF pulse type Normal Gradient mode Normal Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;HE3,4;NE1,2;SP1 Acquisition duration 0 ms Mode Min flip angle Min flip angle 130 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\With Contrast\STIR Sag Right TA:3:47 PAT:2 Voxel size:0.7×0.7×3.0 mm Rel. SNR:1.00 :tir_rr --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 45 Dist. factor 10 % Position R78.0 P21.8 H0.3 mm Orientation S > T3.6 Phase enc. dir. A >> P AutoAlign --- Phase oversampling 40 % FoV read 220 mm FoV phase 100.0 % Slice thickness 3.0 mm TR 3000.0 ms TE 35.0 ms Averages 2 Concatenations 3 Filter Distortion Corr.(2D), Elliptical filter Coil elements BO1;HE3,4;NE1,2;SP1 Contrast TD 0.0 ms MTC Off Magn. preparation Slice-sel. IR TI 220 ms Freeze suppressed tissue Off Flip angle 160 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 320 Phase resolution 75 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 32 Reference scan mode Self-calibration Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 45 Dist. factor 10 % Position R78.0 P21.8 H3.3 mm Phase enc. dir. A >> P Phase oversampling 40 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 Off BO3 Off HE1 Off HE3 On NE1 On HE2 Off HE4 On NE2 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 On SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 3 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation Slice-sel. IR TI 220 ms Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 300 Hz/Px Flow comp. Read Allowed delay 60 s Echo spacing 8.82 ms Define Turbo factor Turbo factor 15 Echo trains per slice 12 RF pulse type Normal Gradient mode Normal Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;HE3,4;NE1,2;SP1 Acquisition duration 0 ms Mode Min flip angle Min flip angle 130 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Brachial Plexus\With Contrast\DTI Ax TA:4:11 PAT:2 Voxel size:2.5×2.5×5.0 mm Rel. SNR:1.00 :epse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 30 Dist. factor 0 % Position R12.9 P29.6 F0.4 mm Orientation Transversal Phase enc. dir. P >> A AutoAlign --- Phase oversampling 0 % FoV read 320 mm FoV phase 81.5 % Slice thickness 5.0 mm TR 4700 ms TE 69.0 ms Averages 2 Concatenations 1 Filter Raw filter Coil elements BO2,3;SP4,5 Contrast MTC Off Magn. preparation None Fat suppr. SPAIR Fat sat. mode Strong Averaging mode Long term Delay in TR 0 ms Reconstruction Magnitude Multiple series Off Resolution Base resolution 130 Phase resolution 100 % Phase partial Fourier Off Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Separate Distortion Corr. Off Prescan Normalize Off Normalize Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Dynamic Field Corr. Off Geometry Nr. of slice groups 1 Slices 30 Dist. factor 0 % Position R12.9 P29.6 F130.4 mm Phase enc. dir. P >> A Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. SPAIR Special sat. None Fat sat. mode Strong Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off BO1 Off BO2 On BO3 On BO1 Off BO2 Off BO3 Off SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal R >> L Coronal A >> P Transversal F >> H Coil Combine Mode Sum of Squares AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R12.9 P29.6 F130.4 mm Rotation -180.00 deg R >> L 320 mm A >> P 261 mm F >> H 150 mm Frequency 1H 123.200000 MHz Correction factor 1 ExtExciteRF 1H 140.463 V Gain High Table position 130 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Averaging mode Long term Multi-slice mode Interleaved Bandwidth 2136 Hz/Px Optimization None Free echo spacing Off Echo spacing 0.59 ms EPI factor 106 RF pulse type Normal Gradient mode Fast* TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO2,3;SP4,5 Acquisition duration 0 ms BOLD Delay in TR 0 ms Diffusion mode MDDW Diff. weightings 3 b-value 1 0 s/mm² Diff. weighted images On Trace weighted images On ADC maps On FA maps On Mosaic On Tensor On Distortion Corr. Off b-Value >= 0 s/mm² Exponential ADC Maps Off Invert Gray Scale Off Calculated Image Off Calculated bValue 1400 s/mm² SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Tarsal Tunnel\Routine\localizer_tra TA:0:12 PAT:Off Voxel size:0.8×0.8×6.0 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments Off Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 5 Dist. factor 50 % Position Isocenter Orientation Transversal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 38 % FoV read 400 mm FoV phase 100.0 % Slice thickness 6.0 mm TR 7.3 ms TE 3.60 ms Averages 1 Concatenations 5 Filter Distortion Corr.(2D), Elliptical filter Coil elements FA;TO Contrast TD 0 ms MTC Off Magn. preparation None Flip angle 20 deg Fat suppr. None Water suppr. None SWI Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 80 % Phase partial Fourier Off Interpolation On PAT mode None Image Filter Off Distortion Corr. On TD 0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 5 Dist. factor 50 % Position Isocenter Phase enc. dir. R >> L Phase oversampling 38 % Multi-slice mode Sequential Series Interleaved Saturation mode Standard Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Tim CT mode Off System Body Off FA On TO On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode REF Table position F Table position 0 mm MSMA S - C - T Sagittal R >> L Coronal A >> P Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Off Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SRFExcit 1H 92.684 V Gain High Table position 0 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Segments 1 Tagging None Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Phase stabilisation Off Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Allowed Contrasts 1 Bandwidth 330 Hz/Px Flow comp. No Allowed delay 0 s RF pulse type Fast Gradient mode Fast Excitation Slice-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements FA;TO Acquisition duration 0 ms Mode Off BOLD Subtract Off Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Tarsal Tunnel\Routine\localizer_sag+cor+tra TA:0:20 PAT:Off Voxel size:0.5×0.5×6.0 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments Off Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 3 Slices 3 Dist. factor 50 % Position L0.0 A65.1 H0.2 mm Orientation Sagittal Phase enc. dir. A >> P AutoAlign --- Phase oversampling 38 % FoV read 280 mm FoV phase 100.0 % Slice thickness 6.0 mm TR 7.3 ms TE 3.60 ms Averages 1 Concatenations 9 Filter Distortion Corr.(2D), Elliptical filter Coil elements FA;TO Contrast TD 0 ms MTC Off Magn. preparation None Flip angle 20 deg Fat suppr. None Water suppr. None SWI Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 80 % Phase partial Fourier Off Interpolation On PAT mode None Image Filter Off Distortion Corr. On TD 0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 3 Slices 3 Dist. factor 50 % Position L0.0 A65.1 H31.2 mm Phase enc. dir. A >> P Phase oversampling 38 % Multi-slice mode Sequential Series Interleaved Saturation mode Standard Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Tim CT mode Off System Body Off FA On TO On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal R >> L Coronal A >> P Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Off Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SRFExcit 1H 92.684 V Gain High Table position 31 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Segments 1 Tagging None Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Phase stabilisation Off Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Allowed Contrasts 1 Bandwidth 330 Hz/Px Flow comp. No Allowed delay 0 s RF pulse type Fast Gradient mode Fast Excitation Slice-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements FA;TO Acquisition duration 0 ms Mode Off BOLD Subtract Off Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Tarsal Tunnel\Routine\T2 Ax fs TA:2:15 PAT:Off Voxel size:0.3×0.3×3.0 mm Rel. SNR:1.00 :tse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments Off Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 30 Dist. factor 20 % Position L12.2 A5.4 H21.0 mm Orientation Transversal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 0 % FoV read 150 mm FoV phase 62.5 % Slice thickness 3.0 mm TR 6340.0 ms TE 65.0 ms Averages 1 Concatenations 1 Filter Distortion Corr.(2D), Normalize, Elliptical filter Coil elements FA;TO Contrast MTC Off Magn. preparation None Flip angle 150 deg Fat suppr. SPAIR Fat sat. mode Strong Water suppr. None Restore magn. Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 512 Phase resolution 80 % Phase partial Fourier Allowed Trajectory Cartesian Interpolation Off PAT mode None Image Filter Off Distortion Corr. On Mode 2D Unfiltered images Off Prescan Normalize Off Normalize On Cut off 0 Width 0 Unfiltered images Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 30 Dist. factor 20 % Position L12.2 A5.4 H52.0 mm Phase enc. dir. R >> L Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. SPAIR Water suppr. None Special sat. Parallel H Gap 10 mm Thickness 60 mm Fat sat. mode Strong Special sat. Parallel H Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off FA On TO On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode REF Table position H Table position 0 mm MSMA S - C - T Sagittal R >> L Coronal A >> P Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position L12.2 A5.4 H52.0 mm Rotation 90.00 deg A >> P 150 mm R >> L 94 mm F >> H 108 mm Frequency 1H 123.200000 MHz Correction factor 1 optfsOptfs 1H 135.123 V Gain High Table position 31 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Short term Multi-slice mode Interleaved Reduce Motion Sens. Off Contrasts 1 Bandwidth 250 Hz/Px Flow comp. No Allowed delay 60 s Echo spacing 10.8 ms Define Turbo factor Turbo factor 14 Echo trains per slice 19 RF pulse type Fast Gradient mode Normal Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements FA;TO Acquisition duration 0 ms Mode Off BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Tarsal Tunnel\Routine\T1 Ax TA:3:44 PAT:Off Voxel size:0.5×0.5×3.0 mm Rel. SNR:1.00 :tse_rs --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments Off Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 30 Dist. factor 20 % Position L10.2 A5.4 H21.0 mm Orientation Transversal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 0 % FoV read 150 mm FoV phase 62.5 % Slice thickness 3.0 mm TR 680.0 ms TE 12.0 ms Averages 2 Concatenations 3 Filter Distortion Corr.(2D), Prescan Normalize, Elliptical filter Coil elements FA;TO Contrast TD 0.0 ms MTC Off Magn. preparation None Flip angle 150 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Off Resolution Base resolution 320 Phase resolution 80 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode None Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 30 Dist. factor 20 % Position L10.2 A5.4 H52.0 mm Phase enc. dir. R >> L Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. Parallel H Gap 10 mm Thickness 60 mm Special sat. Parallel H Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off FA On TO On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode FIX Table position H Table position 0 mm MSMA S - C - T Sagittal R >> L Coronal A >> P Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position L10.2 A5.4 H52.0 mm Rotation 90.00 deg A >> P 150 mm R >> L 94 mm F >> H 108 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 31 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Short term Multi-slice mode Interleaved Reduce Motion Sens. Off Contrasts 1 Bandwidth 230 Hz/Px Flow comp. Slice Allowed delay 30 s Echo spacing 12.4 ms Define Turbo factor Turbo factor 3 Echo trains per slice 54 RF pulse type Normal Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements FA;TO Acquisition duration 0 ms Mode Off BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Tarsal Tunnel\Routine\PD Sag fs TA:4:14 PAT:Off Voxel size:0.4×0.4×3.0 mm Rel. SNR:1.00 :tse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments Off Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 20 Dist. factor 20 % Position L4.5 P0.0 H31.2 mm Orientation S > C-5.0 Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % FoV read 150 mm FoV phase 100.0 % Slice thickness 3.0 mm TR 3000.0 ms TE 40.0 ms Averages 2 Concatenations 1 Filter Distortion Corr.(2D), Prescan Normalize, Elliptical filter Coil elements FA;TO Contrast MTC Off Magn. preparation None Flip angle 160 deg Fat suppr. Fat sat. Fat sat. mode Weak Water suppr. None Restore magn. Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Off Resolution Base resolution 384 Phase resolution 85 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode None Image Filter Off Distortion Corr. On Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 20 Dist. factor 20 % Position L4.5 P0.0 H62.2 mm Phase enc. dir. A >> P Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. Fat sat. Water suppr. None Special sat. None Fat sat. mode Weak Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off FA On TO On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode REF Table position H Table position 0 mm MSMA S - C - T Sagittal R >> L Coronal A >> P Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position L4.5 P0.0 H62.2 mm Rotation 35.60 deg F >> H 150 mm A >> P 150 mm R >> L 72 mm Frequency 1H 123.200000 MHz Correction factor 1 SLoopFCSatNS 1H 83.814 V Gain High Table position 31 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Short term Multi-slice mode Interleaved Reduce Motion Sens. Off Contrasts 1 Bandwidth 246 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 9.99 ms Define Turbo factor Turbo factor 8 Echo trains per slice 41 RF pulse type Fast Gradient mode Normal Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements FA;TO Acquisition duration 0 ms Mode Off BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Tarsal Tunnel\Routine\T2 Coro fs TA:6:34 PAT:Off Voxel size:0.5×0.5×3.0 mm Rel. SNR:1.00 :tse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments Off Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 30 Dist. factor 20 % Position L28.2 A2.1 H28.1 mm Orientation T > C-41.4 > S3.3 Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % FoV read 150 mm FoV phase 100.0 % Slice thickness 3.0 mm TR 5610.0 ms TE 66.0 ms Averages 2 Concatenations 1 Filter Distortion Corr.(2D), Prescan Normalize, Elliptical filter Coil elements FA;TO Contrast MTC Off Magn. preparation None Flip angle 160 deg Fat suppr. SPAIR Fat sat. mode Weak Water suppr. None Restore magn. Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Off Resolution Base resolution 320 Phase resolution 85 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode None Image Filter Off Distortion Corr. On Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 30 Dist. factor 20 % Position L28.2 A2.1 H59.1 mm Phase enc. dir. A >> P Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. SPAIR Water suppr. None Special sat. None Fat sat. mode Weak Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off FA On TO On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode REF Table position H Table position 0 mm MSMA S - C - T Sagittal R >> L Coronal A >> P Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position L28.2 A2.1 H59.1 mm Rotation 35.60 deg R >> L 150 mm A >> P 150 mm F >> H 108 mm Frequency 1H 123.200000 MHz Correction factor 1 optfsOptfs 1H 135.123 V Gain High Table position 31 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Short term Multi-slice mode Interleaved Reduce Motion Sens. Off Contrasts 1 Bandwidth 244 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 9.45 ms Define Turbo factor Turbo factor 8 Echo trains per slice 34 RF pulse type Fast Gradient mode Normal Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements FA;TO Acquisition duration 0 ms Mode Off BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Tarsal Tunnel\Routine\T1 Coro TA:3:42 PAT:Off Voxel size:0.4×0.4×3.0 mm Rel. SNR:1.00 :tse_rs --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments Off Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 30 Dist. factor 20 % Position L6.3 P0.0 H27.2 mm Orientation T > C-41.4 > S3.3 Phase enc. dir. R >> L AutoAlign --- Phase oversampling 0 % FoV read 150 mm FoV phase 62.5 % Slice thickness 3.0 mm TR 856.0 ms TE 14.0 ms Averages 2 Concatenations 2 Filter Distortion Corr.(2D), Prescan Normalize, Elliptical filter Coil elements FA;TO Contrast TD 0.0 ms MTC Off Magn. preparation None Flip angle 150 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Off Resolution Base resolution 384 Phase resolution 80 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode None Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 30 Dist. factor 20 % Position L6.3 P0.0 H58.2 mm Phase enc. dir. R >> L Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. Parallel H Gap 10 mm Thickness 60 mm Special sat. Parallel H Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off FA On TO On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode FIX Table position H Table position 0 mm MSMA S - C - T Sagittal R >> L Coronal A >> P Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position L6.3 P0.0 H58.2 mm Rotation 125.60 deg A >> P 150 mm R >> L 94 mm F >> H 108 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 31 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Short term Multi-slice mode Interleaved Reduce Motion Sens. Off Contrasts 1 Bandwidth 228 Hz/Px Flow comp. Slice Allowed delay 30 s Echo spacing 13.6 ms Define Turbo factor Turbo factor 3 Echo trains per slice 64 RF pulse type Normal Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements FA;TO Acquisition duration 0 ms Mode Off BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Tarsal Tunnel\Routine\Vibe Coro fs TA:1:54 PAT:Off Voxel size:0.8×0.8×0.8 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments Off Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position L9.5 P6.5 H35.2 mm Orientation C > S5.0 Phase enc. dir. F >> H AutoAlign --- Phase oversampling 0 % Slice oversampling 25.0 % FoV read 150 mm FoV phase 68.8 % Slice thickness 0.80 mm TR 11.00 ms TE 4.06 ms Averages 1 Concatenations 1 Filter Distortion Corr.(2D), Normalize Coil elements FA;TO Contrast Flip angle 10.0 deg Fat suppr. Water excit. normal Water suppr. None Dixon No Dixon Save original images On Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Off Resolution Base resolution 192 Phase resolution 90 % Phase partial Fourier 6/8 Interpolation Off PAT mode None Image Filter Off Distortion Corr. On Mode 2D Unfiltered images Off Prescan Normalize Off Normalize On Cut off 0 Width 0 Unfiltered images Off B1 filter Off Raw filter Off Elliptical filter Off POCS Off Slice resolution 83 % Slice partial Fourier 7/8 Geometry Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position L9.5 P6.5 H66.2 mm Phase enc. dir. F >> H Phase oversampling 0 % Slice oversampling 25.0 % Slices per slab 128 Multi-slice mode Sequential Series Ascending Nr. of sat. regions 0 Position mode L-P-H Fat suppr. Water excit. normal Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off FA On TO On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 Off Position mode L-P-H Positioning mode FIX Table position H Table position 0 mm MSMA S - C - T Sagittal R >> L Coronal A >> P Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Off Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position L9.5 P6.5 H66.2 mm Rotation 125.60 deg R >> L 150 mm F >> H 104 mm A >> P 103 mm Frequency 1H 123.200000 MHz Correction factor 1 ExcitWEs 0 1H 51.370 V Gain Low Table position 31 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning On Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Weak Contrasts 1 Bandwidth 190 Hz/Px Optimization Min. TE Allowed delay 0 s RF pulse type Fast Gradient mode Fast Excitation Slab-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements FA;TO Acquisition duration 0 ms Mode Off BOLD 3D centric reordering Off Time to center 52.8 s Subtract Off Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Tarsal Tunnel\Routine\PSIF Coro Obl b80 TA:6:53 PAT:Off Voxel size:0.8×0.8×0.9 mm Rel. SNR:1.00 :ps_rr --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments Off Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position Isocenter Orientation Coronal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 0 % Slice oversampling 0.0 % FoV read 160 mm FoV phase 100.0 % Slice thickness 0.90 mm TR 10.00 ms TE 2.57 ms Averages 1 Filter None Coil elements FA;TO Contrast Flip angle 35 deg Fat suppr. Water excit. normal Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 192 Phase resolution 100 % Phase partial Fourier Off Interpolation Off PAT mode None Image Filter Off Distortion Corr. Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Slice resolution 88 % Slice partial Fourier Off Geometry Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position Isocenter Phase enc. dir. R >> L Phase oversampling 0 % Slice oversampling 0.0 % Slices per slab 288 Multi-slice mode Sequential Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. Water excit. normal Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off FA On TO On Position mode L-P-H Positioning mode REF Table position H Table position 0 mm MSMA S - C - T Sagittal R >> L Coronal A >> P Transversal F >> H Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg F >> H 160 mm R >> L 160 mm A >> P 260 mm Frequency 1H 123.200000 MHz Correction factor 1 ExcitWEs 0 1H 408.000 V Gain Low Table position 0 mm Img. Scale. Cor. 1.000 Physio Segments 1 Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning On Averaging mode Long term Multi-slice mode Sequential Asymmetric echo Off Bandwidth 521 Hz/Px Flow comp. Read Optimization Min. TE RF pulse type Normal Gradient mode Fast Excitation Slab-sel. TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements FA;TO Acquisition duration 0 ms Mode Off BOLD Diffusion mode Phase Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. Off Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\Routine\SCOUT TA:0:35 PAT:Off Voxel size:0.7×0.7×6.0 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 3 Slices 5 Dist. factor 50 % Position R18.9 P33.1 H0.4 mm Orientation Coronal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 38 % FoV read 380 mm FoV phase 100.0 % Slice thickness 6.0 mm TR 7.8 ms TE 3.69 ms Averages 1 Concatenations 15 Filter Distortion Corr.(2D) Coil elements BO1-3;SP2-5 Contrast TD 0 ms MTC Off Magn. preparation None Flip angle 20 deg Fat suppr. None Water suppr. None SWI Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 80 % Phase partial Fourier Off Interpolation On PAT mode None Image Filter Off Distortion Corr. On TD 0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Geometry Nr. of slice groups 3 Slices 5 Dist. factor 50 % Position R18.9 P33.1 H114.4 mm Phase enc. dir. R >> L Phase oversampling 38 % Multi-slice mode Sequential Series Interleaved Saturation mode Standard Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Tim CT mode Off System Body Off BO1 On BO2 On BO3 On BO1 Off BO2 Off BO3 Off SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 On SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SRFExcit 1H 53.909 V Gain High Table position 114 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Segments 1 Tagging None Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Phase stabilisation Off Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Off Contrasts 1 Bandwidth 320 Hz/Px Flow comp. No Allowed delay 0 s RF pulse type Normal Gradient mode Normal Excitation Slice-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1-3;SP2-5 Acquisition duration 0 ms Mode Off BOLD Subtract Off Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\Routine\SCOUT TA:0:35 PAT:Off Voxel size:0.7×0.7×6.0 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 3 Slices 5 Dist. factor 50 % Position R18.9 P25.8 F0.1 mm Orientation Coronal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 38 % FoV read 380 mm FoV phase 100.0 % Slice thickness 6.0 mm TR 7.8 ms TE 3.69 ms Averages 1 Concatenations 15 Filter Distortion Corr.(2D) Coil elements BO1-3;BO2,3;SP2-5 Contrast TD 0 ms MTC Off Magn. preparation None Flip angle 20 deg Fat suppr. None Water suppr. None SWI Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 80 % Phase partial Fourier Off Interpolation On PAT mode None Image Filter Off Distortion Corr. On TD 0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Geometry Nr. of slice groups 3 Slices 5 Dist. factor 50 % Position R18.9 P25.8 F65.1 mm Phase enc. dir. R >> L Phase oversampling 38 % Multi-slice mode Sequential Series Interleaved Saturation mode Standard Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Tim CT mode Off System Body Off BO1 On BO2 On BO3 On BO1 Off BO2 On BO3 On SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 On SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SRFExcit 1H 53.909 V Gain High Table position 65 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Segments 1 Tagging None Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Phase stabilisation Off Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Off Contrasts 1 Bandwidth 320 Hz/Px Flow comp. No Allowed delay 0 s RF pulse type Normal Gradient mode Normal Excitation Slice-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1-3;BO2,3;SP2-5 Acquisition duration 0 ms Mode Off BOLD Subtract Off Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\Routine\STIR Sag BILATERAL TA:2:46 PAT:2 Voxel size:0.7×0.7×3.0 mm Rel. SNR:1.00 :tir_rr --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 40 Dist. factor 33 % Position R15.6 P20.8 F0.1 mm Orientation S > T-1.2 > C0.6 Phase enc. dir. A >> P AutoAlign --- Phase oversampling 25 % FoV read 220 mm FoV phase 100.0 % Slice thickness 3.0 mm TR 3922.2 ms TE 46.0 ms Averages 2 Concatenations 2 Filter Distortion Corr.(2D), Prescan Normalize, Elliptical filter Coil elements BO1;BO2,3;SP3,4 Contrast TD 0.0 ms MTC Off Magn. preparation Slice-sel. IR TI 220 ms Freeze suppressed tissue Off Flip angle 150 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 320 Phase resolution 75 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 32 Reference scan mode Self-calibration Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 40 Dist. factor 33 % Position R15.6 P20.8 F6.1 mm Phase enc. dir. A >> P Phase oversampling 25 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 Off BO3 Off BO1 Off BO2 On BO3 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 6 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation Slice-sel. IR TI 220 ms Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 260 Hz/Px Flow comp. Read Allowed delay 60 s Echo spacing 9.22 ms Define Turbo factor Turbo factor 16 Echo trains per slice 10 RF pulse type Normal Gradient mode Normal Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;BO2,3;SP3,4 Acquisition duration 0 ms Mode Min flip angle Min flip angle 130 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\Routine\T1 Sag BILATERAL TA:4:23 PAT:2 Voxel size:0.6×0.6×3.0 mm Rel. SNR:1.00 :tse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 40 Dist. factor 33 % Position R15.6 P20.8 F0.1 mm Orientation S > T-1.2 > C0.6 Phase enc. dir. A >> P AutoAlign --- Phase oversampling 25 % FoV read 220 mm FoV phase 100.0 % Slice thickness 3.0 mm TR 800.0 ms TE 9.9 ms Averages 3 Concatenations 3 Filter Distortion Corr.(2D), Prescan Normalize, Elliptical filter Coil elements BO1;BO2,3;SP3,4 Contrast TD 0.0 ms MTC Off Magn. preparation None Flip angle 150 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 384 Phase resolution 75 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 32 Reference scan mode Self-calibration Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 40 Dist. factor 33 % Position R15.6 P20.8 F6.1 mm Phase enc. dir. A >> P Phase oversampling 25 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 Off BO3 Off BO1 Off BO2 On BO3 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 VExcit 1H 152.608 V Gain High Table position 6 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 250 Hz/Px Flow comp. No Allowed delay 60 s Echo spacing 9.86 ms Define Turbo factor Turbo factor 5 Echo trains per slice 36 RF pulse type Low SAR Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;BO2,3;SP3,4 Acquisition duration 0 ms Mode TR Max. TR 750.0 ms BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\Routine\t2_spc_stir_cor_p2_iso TA:3:21 PAT:3 Voxel size:1.0×1.0×1.0 mm Rel. SNR:1.00 :spcir --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Position R17.0 P21.2 H0.1 mm Orientation Coronal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 20 % Slice oversampling 0.0 % FoV read 380 mm FoV phase 100.0 % Slice thickness 1.00 mm TR 1500 ms TE 112.0 ms Concatenations 1 Filter Raw filter, Distortion Corr.(2D), Prescan Normalize Coil elements BO1,2;BO2,3;SP2-5 Contrast MTC Off Magn. preparation Non-sel. IR TI 220 ms Fat suppr. None Water suppr. None Restore magn. Off Measurements 1 Reconstruction Magnitude Multiple series Each measurement Gradient 1st moment in read 300 mT ms^2/m Gradient 1st moment in phase 300 mT ms^2/m Gradient 1st moment in slice 0 mT ms^2/m Resolution Base resolution 384 Phase resolution 100 % Phase partial Fourier Allowed Interpolation Off PAT mode GRAPPA Accel. factor PE 3 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Accel. factor 3D 1 Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Slice resolution 50 % Slice partial Fourier 6/8 Geometry Nr. of slab groups 1 Slabs 1 Position R17.0 P21.2 F50.9 mm Phase enc. dir. R >> L Phase oversampling 20 % Slice oversampling 0.0 % Slices per slab 192 Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off System Body Off BO1 On BO2 On BO3 Off BO1 Off BO2 On BO3 On SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 On SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SLoopIRns1 1H 408.000 V ! Gain High Table position 51 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Trigger delay 0 ms Magn. preparation Non-sel. IR TI 220 ms Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning Off Reordering Linear Bandwidth 814 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 3.33 ms Adiabatic-mode Off Turbo factor 128 Echo train duration 310 RF pulse type Fast Gradient mode Fast Excitation Slab-sel. Flip angle mode T2 var TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;BO2,3;SP2-5 Acquisition duration 0 ms Organ under exam. None BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\Routine\T2spc Sag (lumbar) TA:7:26 PAT:2 Voxel size:0.7×0.7×0.6 mm Rel. SNR:1.00 :spcR --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Position R21.2 P18.5 F0.3 mm Orientation S > T-1.6 Phase enc. dir. H >> F AutoAlign --- Phase oversampling 80 % Slice oversampling 0.0 % FoV read 250 mm FoV phase 100.0 % Slice thickness 0.60 mm TR 1500 ms TE 136.0 ms Concatenations 1 Filter Raw filter, Distortion Corr.(2D), Prescan Normalize Coil elements BO1;BO1-3;SP2-4 Contrast MTC Off Magn. preparation None Flip angle 120 deg Fat suppr. None Water suppr. None Restore magn. On Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 384 Phase resolution 100 % Phase partial Fourier Allowed Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Accel. factor 3D 1 Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Slice resolution 100 % Slice partial Fourier 6/8 Geometry Nr. of slab groups 1 Slabs 1 Position R21.2 P18.5 H24.7 mm Phase enc. dir. H >> F Phase oversampling 80 % Slice oversampling 0.0 % Slices per slab 64 Series Ascending Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. On System Body Off BO1 On BO2 Off BO3 Off BO1 On BO2 On BO3 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 On SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SRFExcit 1H 395.248 V ! Gain High Table position 25 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Trigger delay 0 ms Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning Off Reordering Linear Bandwidth 383 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 5.61 ms Adiabatic-mode Off Turbo factor 80 Echo train duration 353 RF pulse type Normal Gradient mode Fast Excitation Slab-sel. Flip angle mode Constant TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;BO1-3;SP2-4 Acquisition duration 0 ms Organ under exam. None BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\Routine\T1 Ax TA:3:14 PAT:4 Voxel size:0.8×0.8×4.0 mm Rel. SNR:1.00 :tse 1 / 2 --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 30 Dist. factor 25 % Position R20.5 P23.7 H0.3 mm Orientation Transversal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 100 % FoV read 250 mm FoV phase 100.0 % Slice thickness 4.0 mm TR 650.0 ms TE 9.0 ms Averages 3 Concatenations 2 Filter Distortion Corr.(2D), Prescan Normalize Coil elements BO1;BO2,3;SP3,4 Contrast TD 0.0 ms MTC Off Magn. preparation None Flip angle 180 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Off Resolution Base resolution 320 Phase resolution 80 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 4 Ref. lines PE 26 Reference scan mode Integrated Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Geometry Nr. of slice groups 1 Slices 30 Dist. factor 25 % Position R20.5 P23.7 F4.7 mm Phase enc. dir. R >> L Phase oversampling 100 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol On Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 Off BO3 Off BO1 Off BO2 On BO3 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 5 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 381 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 9.04 ms Define Turbo factor Turbo factor 3 Echo trains per slice 49 RF pulse type Normal Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;BO2,3;SP3,4 Acquisition duration 0 ms Mode Min flip angle Min flip angle 150 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\Routine\T1 Ax TA:2:10 PAT:4 Voxel size:0.8×0.8×4.0 mm Rel. SNR:1.00 :tse 2 / 2 --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 25 Dist. factor 25 % Position R20.5 P23.7 F0.2 mm Orientation Transversal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 100 % FoV read 250 mm FoV phase 100.0 % Slice thickness 4.0 mm TR 866.0 ms TE 9.0 ms Averages 3 Concatenations 1 Filter Distortion Corr.(2D), Prescan Normalize Coil elements BO1,2;SP4,5 Contrast MTC Off Magn. preparation None Flip angle 180 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Off Resolution Base resolution 320 Phase resolution 80 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 4 Ref. lines PE 26 Reference scan mode Integrated Image Filter Off Distortion Corr. On Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Geometry Nr. of slice groups 1 Slices 25 Dist. factor 25 % Position R20.5 P23.7 F142.2 mm Phase enc. dir. R >> L Phase oversampling 100 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol On Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 On BO3 Off BO1 Off BO2 Off BO3 Off SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 142 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 381 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 9.04 ms Define Turbo factor Turbo factor 3 Echo trains per slice 49 RF pulse type Normal Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;SP4,5 Acquisition duration 0 ms Mode Min flip angle Min flip angle 150 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\Routine\T2 Ax SPAIR TA:3:32 PAT:4 Voxel size:0.8×0.8×4.0 mm Rel. SNR:1.00 :tse 1 / 2 --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 30 Dist. factor 25 % Position R20.5 P23.7 H0.3 mm Orientation Transversal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 100 % FoV read 250 mm FoV phase 100.0 % Slice thickness 4.0 mm TR 5550.0 ms TE 58.0 ms Averages 3 Concatenations 2 Filter Distortion Corr.(2D), Prescan Normalize Coil elements BO1;BO2,3;SP3,4 Contrast TD 0.0 ms MTC Off Magn. preparation None Flip angle 160 deg Fat suppr. SPAIR Fat sat. mode Weak Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 320 Phase resolution 78 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 4 Ref. lines PE 34 Reference scan mode Integrated Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Geometry Nr. of slice groups 1 Slices 30 Dist. factor 25 % Position R20.5 P23.7 F4.7 mm Phase enc. dir. R >> L Phase oversampling 100 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. SPAIR Water suppr. None Special sat. None Fat sat. mode Weak Special sat. None Set-n-Go Protocol On Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 Off BO3 Off BO1 Off BO2 On BO3 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R20.5 P23.7 F4.7 mm Rotation 87.00 deg A >> P 250 mm R >> L 250 mm F >> H 149 mm Frequency 1H 123.200000 MHz Correction factor 1 optfsOptfs 1H 135.123 V Gain High Table position 5 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 300 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 9.68 ms Define Turbo factor Turbo factor 25 Echo trains per slice 6 RF pulse type Low SAR Gradient mode Normal Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;BO2,3;SP3,4 Acquisition duration 0 ms Mode Min flip angle Min flip angle 150 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\Routine\T2 Ax SPAIR TA:3:34 PAT:4 Voxel size:0.8×0.8×4.0 mm Rel. SNR:1.00 :tse 2 / 2 --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 25 Dist. factor 25 % Position R20.5 P23.7 F0.2 mm Orientation Transversal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 100 % FoV read 250 mm FoV phase 100.0 % Slice thickness 4.0 mm TR 4820.0 ms TE 58.0 ms Averages 3 Concatenations 2 Filter Distortion Corr.(2D), Prescan Normalize Coil elements BO1,2;SP4,5 Contrast TD 0.0 ms MTC Off Magn. preparation None Flip angle 160 deg Fat suppr. SPAIR Fat sat. mode Weak Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 320 Phase resolution 82 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 4 Ref. lines PE 50 Reference scan mode Integrated Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Geometry Nr. of slice groups 1 Slices 25 Dist. factor 25 % Position R20.5 P23.7 F142.2 mm Phase enc. dir. R >> L Phase oversampling 100 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. SPAIR Water suppr. None Special sat. None Fat sat. mode Weak Special sat. None Set-n-Go Protocol On Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 On BO3 Off BO1 Off BO2 Off BO3 Off SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R20.5 P23.7 F142.2 mm Rotation 87.00 deg A >> P 250 mm R >> L 250 mm F >> H 124 mm Frequency 1H 123.200000 MHz Correction factor 1 optfsOptfs 1H 135.123 V Gain High Table position 142 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 300 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 9.68 ms Define Turbo factor Turbo factor 24 Echo trains per slice 7 RF pulse type Low SAR Gradient mode Normal Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;SP4,5 Acquisition duration 0 ms Mode Min flip angle Min flip angle 150 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\Routine\ep2d_dti_12 direction_p2 TA:4:11 PAT:2 Voxel size:2.5×2.5×5.0 mm Rel. SNR:1.00 :epse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 30 Dist. factor 0 % Position R15.2 P29.6 F0.5 mm Orientation Transversal Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % FoV read 320 mm FoV phase 81.5 % Slice thickness 5.0 mm TR 4700 ms TE 69.0 ms Averages 2 Concatenations 1 Filter Raw filter Coil elements BO1,2;BO3;SP4,5 Contrast MTC Off Magn. preparation None Fat suppr. SPAIR Fat sat. mode Strong Averaging mode Long term Delay in TR 0 ms Reconstruction Magnitude Multiple series Off Resolution Base resolution 130 Phase resolution 100 % Phase partial Fourier Off Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Separate Distortion Corr. Off Prescan Normalize Off Normalize Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Dynamic Field Corr. Off Geometry Nr. of slice groups 1 Slices 30 Dist. factor 0 % Position R15.2 P29.6 F117.5 mm Phase enc. dir. A >> P Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. SPAIR Special sat. None Fat sat. mode Strong Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off BO1 On BO2 On BO3 Off BO1 Off BO2 Off BO3 On SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal R >> L Coronal A >> P Transversal F >> H Coil Combine Mode Sum of Squares AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V ! Position R15.2 P29.6 F117.5 mm ! Rotation -3.20 deg ! R >> L 320 mm ! A >> P 197 mm ! F >> H 150 mm Frequency 1H 123.200000 MHz Correction factor 1 ExtExciteRF 1H 140.463 V Gain High Table position 117 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Averaging mode Long term Multi-slice mode Interleaved Bandwidth 2136 Hz/Px Optimization None Free echo spacing Off Echo spacing 0.59 ms EPI factor 106 RF pulse type Normal Gradient mode Fast* TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;BO3;SP4,5 Acquisition duration 0 ms BOLD Delay in TR 0 ms Diffusion mode MDDW Diff. weightings 3 b-value 1 0 s/mm² Diff. weighted images On Trace weighted images On ADC maps On FA maps On Mosaic On Tensor On Distortion Corr. Off b-Value >= 0 s/mm² Exponential ADC Maps Off Invert Gray Scale Off Calculated Image Off Calculated bValue 1400 s/mm² SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\With Contrast\SCOUT TA:0:35 PAT:Off Voxel size:0.7×0.7×6.0 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 3 Slices 5 Dist. factor 50 % Position R18.9 P33.1 H0.4 mm Orientation Coronal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 38 % FoV read 380 mm FoV phase 100.0 % Slice thickness 6.0 mm TR 7.8 ms TE 3.69 ms Averages 1 Concatenations 15 Filter Distortion Corr.(2D) Coil elements BO1-3;SP2-5 Contrast TD 0 ms MTC Off Magn. preparation None Flip angle 20 deg Fat suppr. None Water suppr. None SWI Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 80 % Phase partial Fourier Off Interpolation On PAT mode None Image Filter Off Distortion Corr. On TD 0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Geometry Nr. of slice groups 3 Slices 5 Dist. factor 50 % Position R18.9 P33.1 H114.4 mm Phase enc. dir. R >> L Phase oversampling 38 % Multi-slice mode Sequential Series Interleaved Saturation mode Standard Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Tim CT mode Off System Body Off BO1 On BO2 On BO3 On BO1 Off BO2 Off BO3 Off SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 On SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SRFExcit 1H 53.909 V Gain High Table position 114 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Segments 1 Tagging None Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Phase stabilisation Off Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Off Contrasts 1 Bandwidth 320 Hz/Px Flow comp. No Allowed delay 0 s RF pulse type Normal Gradient mode Normal Excitation Slice-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1-3;SP2-5 Acquisition duration 0 ms Mode Off BOLD Subtract Off Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\With Contrast\SCOUT TA:0:35 PAT:Off Voxel size:0.7×0.7×6.0 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments On Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 3 Slices 5 Dist. factor 50 % Position R18.9 P25.8 F0.1 mm Orientation Coronal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 38 % FoV read 380 mm FoV phase 100.0 % Slice thickness 6.0 mm TR 7.8 ms TE 3.69 ms Averages 1 Concatenations 15 Filter Distortion Corr.(2D) Coil elements BO1-3;BO2,3;SP2-5 Contrast TD 0 ms MTC Off Magn. preparation None Flip angle 20 deg Fat suppr. None Water suppr. None SWI Off Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 256 Phase resolution 80 % Phase partial Fourier Off Interpolation On PAT mode None Image Filter Off Distortion Corr. On TD 0 ms Mode 2D Unfiltered images Off Prescan Normalize Off Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Geometry Nr. of slice groups 3 Slices 5 Dist. factor 50 % Position R18.9 P25.8 F65.1 mm Phase enc. dir. R >> L Phase oversampling 38 % Multi-slice mode Sequential Series Interleaved Saturation mode Standard Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Tim CT mode Off System Body Off BO1 On BO2 On BO3 On BO1 Off BO2 On BO3 On SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 On SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SRFExcit 1H 53.909 V Gain High Table position 65 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Segments 1 Tagging None Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Phase stabilisation Off Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Off Contrasts 1 Bandwidth 320 Hz/Px Flow comp. No Allowed delay 0 s RF pulse type Normal Gradient mode Normal Excitation Slice-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1-3;BO2,3;SP2-5 Acquisition duration 0 ms Mode Off BOLD Subtract Off Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\With Contrast\STIR Sag BILATERAL TA:2:46 PAT:2 Voxel size:0.7×0.7×3.0 mm Rel. SNR:1.00 :tir_rr --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 40 Dist. factor 33 % Position R15.6 P20.8 F0.1 mm Orientation S > T-1.2 > C0.6 Phase enc. dir. A >> P AutoAlign --- Phase oversampling 25 % FoV read 220 mm FoV phase 100.0 % Slice thickness 3.0 mm TR 3922.2 ms TE 46.0 ms Averages 2 Concatenations 2 Filter Distortion Corr.(2D), Prescan Normalize, Elliptical filter Coil elements BO1;BO2,3;SP3,4 Contrast TD 0.0 ms MTC Off Magn. preparation Slice-sel. IR TI 220 ms Freeze suppressed tissue Off Flip angle 150 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 320 Phase resolution 75 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 32 Reference scan mode Self-calibration Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 40 Dist. factor 33 % Position R15.6 P20.8 F6.1 mm Phase enc. dir. A >> P Phase oversampling 25 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 Off BO3 Off BO1 Off BO2 On BO3 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 6 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation Slice-sel. IR TI 220 ms Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 260 Hz/Px Flow comp. Read Allowed delay 60 s Echo spacing 9.22 ms Define Turbo factor Turbo factor 16 Echo trains per slice 10 RF pulse type Normal Gradient mode Normal Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;BO2,3;SP3,4 Acquisition duration 0 ms Mode Min flip angle Min flip angle 130 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\With Contrast\T1 Sag BILATERAL TA:4:23 PAT:2 Voxel size:0.6×0.6×3.0 mm Rel. SNR:1.00 :tse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 40 Dist. factor 33 % Position R15.6 P20.8 F0.1 mm Orientation S > T-1.2 > C0.6 Phase enc. dir. A >> P AutoAlign --- Phase oversampling 25 % FoV read 220 mm FoV phase 100.0 % Slice thickness 3.0 mm TR 800.0 ms TE 9.9 ms Averages 3 Concatenations 3 Filter Distortion Corr.(2D), Prescan Normalize, Elliptical filter Coil elements BO1;BO2,3;SP3,4 Contrast TD 0.0 ms MTC Off Magn. preparation None Flip angle 150 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 384 Phase resolution 75 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 32 Reference scan mode Self-calibration Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter On Mode Inplane Geometry Nr. of slice groups 1 Slices 40 Dist. factor 33 % Position R15.6 P20.8 F6.1 mm Phase enc. dir. A >> P Phase oversampling 25 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 Off BO3 Off BO1 Off BO2 On BO3 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 VExcit 1H 152.608 V Gain High Table position 6 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 250 Hz/Px Flow comp. No Allowed delay 60 s Echo spacing 9.86 ms Define Turbo factor Turbo factor 5 Echo trains per slice 36 RF pulse type Low SAR Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;BO2,3;SP3,4 Acquisition duration 0 ms Mode TR Max. TR 750.0 ms BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\With Contrast\t2_spc_stir_cor_p2_iso TA:3:21 PAT:3 Voxel size:1.0×1.0×1.0 mm Rel. SNR:1.00 :spcir --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Position R17.0 P21.2 H0.1 mm Orientation Coronal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 20 % Slice oversampling 0.0 % FoV read 380 mm FoV phase 100.0 % Slice thickness 1.00 mm TR 1500 ms TE 112.0 ms Concatenations 1 Filter Raw filter, Distortion Corr.(2D), Prescan Normalize Coil elements BO1,2;BO2,3;SP2-5 Contrast MTC Off Magn. preparation Non-sel. IR TI 220 ms Fat suppr. None Water suppr. None Restore magn. Off Measurements 1 Reconstruction Magnitude Multiple series Each measurement Gradient 1st moment in read 300 mT ms^2/m Gradient 1st moment in phase 300 mT ms^2/m Gradient 1st moment in slice 0 mT ms^2/m Resolution Base resolution 384 Phase resolution 100 % Phase partial Fourier Allowed Interpolation Off PAT mode GRAPPA Accel. factor PE 3 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Accel. factor 3D 1 Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Slice resolution 50 % Slice partial Fourier 6/8 Geometry Nr. of slab groups 1 Slabs 1 Position R17.0 P21.2 F50.9 mm Phase enc. dir. R >> L Phase oversampling 20 % Slice oversampling 0.0 % Slices per slab 192 Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. Off System Body Off BO1 On BO2 On BO3 Off BO1 Off BO2 On BO3 On SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 On SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SLoopIRns1 1H 408.000 V ! Gain High Table position 51 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Trigger delay 0 ms Magn. preparation Non-sel. IR TI 220 ms Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning Off Reordering Linear Bandwidth 814 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 3.33 ms Adiabatic-mode Off Turbo factor 128 Echo train duration 310 RF pulse type Fast Gradient mode Fast Excitation Slab-sel. Flip angle mode T2 var TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;BO2,3;SP2-5 Acquisition duration 0 ms Organ under exam. None BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\With Contrast\T2spc Sag (lumbar) TA:7:26 PAT:2 Voxel size:0.7×0.7×0.6 mm Rel. SNR:1.00 :spcR --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Position R21.2 P18.5 F0.3 mm Orientation S > T-1.6 Phase enc. dir. H >> F AutoAlign --- Phase oversampling 80 % Slice oversampling 0.0 % FoV read 250 mm FoV phase 100.0 % Slice thickness 0.60 mm TR 1500 ms TE 136.0 ms Concatenations 1 Filter Raw filter, Distortion Corr.(2D), Prescan Normalize Coil elements BO1;BO1-3;SP2-4 Contrast MTC Off Magn. preparation None Flip angle 120 deg Fat suppr. None Water suppr. None Restore magn. On Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 384 Phase resolution 100 % Phase partial Fourier Allowed Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Integrated Image Filter Off Distortion Corr. On Accel. factor 3D 1 Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Slice resolution 100 % Slice partial Fourier 6/8 Geometry Nr. of slab groups 1 Slabs 1 Position R21.2 P18.5 H24.7 mm Phase enc. dir. H >> F Phase oversampling 80 % Slice oversampling 0.0 % Slices per slab 64 Series Ascending Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off Restore magn. On System Body Off BO1 On BO2 Off BO3 Off BO1 On BO2 On BO3 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 On SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position H Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 SRFExcit 1H 395.248 V ! Gain High Table position 25 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Trigger delay 0 ms Magn. preparation None Dark blood Off Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 3D Elliptical scanning Off Reordering Linear Bandwidth 383 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 5.61 ms Adiabatic-mode Off Turbo factor 80 Echo train duration 353 RF pulse type Normal Gradient mode Fast Excitation Slab-sel. Flip angle mode Constant TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;BO1-3;SP2-4 Acquisition duration 0 ms Organ under exam. None BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\With Contrast\T1 Ax TA:3:14 PAT:4 Voxel size:0.8×0.8×4.0 mm Rel. SNR:1.00 :tse 1 / 2 --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 30 Dist. factor 25 % Position R20.5 P23.7 H0.3 mm Orientation Transversal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 100 % FoV read 250 mm FoV phase 100.0 % Slice thickness 4.0 mm TR 650.0 ms TE 9.0 ms Averages 3 Concatenations 2 Filter Distortion Corr.(2D), Prescan Normalize Coil elements BO1;BO2,3;SP3,4 Contrast TD 0.0 ms MTC Off Magn. preparation None Flip angle 180 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Off Resolution Base resolution 320 Phase resolution 80 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 4 Ref. lines PE 26 Reference scan mode Integrated Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Geometry Nr. of slice groups 1 Slices 30 Dist. factor 25 % Position R20.5 P23.7 F4.7 mm Phase enc. dir. R >> L Phase oversampling 100 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol On Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 Off BO3 Off BO1 Off BO2 On BO3 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 5 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 381 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 9.04 ms Define Turbo factor Turbo factor 3 Echo trains per slice 49 RF pulse type Normal Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;BO2,3;SP3,4 Acquisition duration 0 ms Mode Min flip angle Min flip angle 150 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\With Contrast\T1 Ax TA:2:10 PAT:4 Voxel size:0.8×0.8×4.0 mm Rel. SNR:1.00 :tse 2 / 2 --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 25 Dist. factor 25 % Position R20.5 P23.7 F0.2 mm Orientation Transversal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 100 % FoV read 250 mm FoV phase 100.0 % Slice thickness 4.0 mm TR 866.0 ms TE 9.0 ms Averages 3 Concatenations 1 Filter Distortion Corr.(2D), Prescan Normalize Coil elements BO1,2;SP4,5 Contrast MTC Off Magn. preparation None Flip angle 180 deg Fat suppr. None Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Off Resolution Base resolution 320 Phase resolution 80 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 4 Ref. lines PE 26 Reference scan mode Integrated Image Filter Off Distortion Corr. On Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Geometry Nr. of slice groups 1 Slices 25 Dist. factor 25 % Position R20.5 P23.7 F142.2 mm Phase enc. dir. R >> L Phase oversampling 100 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. None Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol On Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 On BO3 Off BO1 Off BO2 Off BO3 Off SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Tune up Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position Isocenter Rotation 0.00 deg R >> L 350 mm A >> P 263 mm F >> H 350 mm Frequency 1H 123.200000 MHz Correction factor 1 Excit 1H 209.456 V Gain High Table position 142 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 381 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 9.04 ms Define Turbo factor Turbo factor 3 Echo trains per slice 49 RF pulse type Normal Gradient mode Fast Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;SP4,5 Acquisition duration 0 ms Mode Min flip angle Min flip angle 150 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\With Contrast\T2 Ax SPAIR TA:3:32 PAT:4 Voxel size:0.8×0.8×4.0 mm Rel. SNR:1.00 :tse 1 / 2 --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 30 Dist. factor 25 % Position R20.5 P23.7 H0.3 mm Orientation Transversal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 100 % FoV read 250 mm FoV phase 100.0 % Slice thickness 4.0 mm TR 5550.0 ms TE 58.0 ms Averages 3 Concatenations 2 Filter Distortion Corr.(2D), Prescan Normalize Coil elements BO1;BO2,3;SP3,4 Contrast TD 0.0 ms MTC Off Magn. preparation None Flip angle 160 deg Fat suppr. SPAIR Fat sat. mode Weak Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 320 Phase resolution 78 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 4 Ref. lines PE 34 Reference scan mode Integrated Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Geometry Nr. of slice groups 1 Slices 30 Dist. factor 25 % Position R20.5 P23.7 F4.7 mm Phase enc. dir. R >> L Phase oversampling 100 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. SPAIR Water suppr. None Special sat. None Fat sat. mode Weak Special sat. None Set-n-Go Protocol On Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 Off BO3 Off BO1 Off BO2 On BO3 On SP5 Off SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R20.5 P23.7 F4.7 mm Rotation 87.00 deg A >> P 250 mm R >> L 250 mm F >> H 149 mm Frequency 1H 123.200000 MHz Correction factor 1 optfsOptfs 1H 135.123 V Gain High Table position 5 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 300 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 9.68 ms Define Turbo factor Turbo factor 25 Echo trains per slice 6 RF pulse type Low SAR Gradient mode Normal Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1;BO2,3;SP3,4 Acquisition duration 0 ms Mode Min flip angle Min flip angle 150 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\With Contrast\T2 Ax SPAIR TA:3:34 PAT:4 Voxel size:0.8×0.8×4.0 mm Rel. SNR:1.00 :tse 2 / 2 --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 25 Dist. factor 25 % Position R20.5 P23.7 F0.2 mm Orientation Transversal Phase enc. dir. R >> L AutoAlign --- Phase oversampling 100 % FoV read 250 mm FoV phase 100.0 % Slice thickness 4.0 mm TR 4820.0 ms TE 58.0 ms Averages 3 Concatenations 2 Filter Distortion Corr.(2D), Prescan Normalize Coil elements BO1,2;SP4,5 Contrast TD 0.0 ms MTC Off Magn. preparation None Flip angle 160 deg Fat suppr. SPAIR Fat sat. mode Weak Water suppr. None Restore magn. Off Averaging mode Long term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 320 Phase resolution 82 % Phase partial Fourier Off Trajectory Cartesian Interpolation Off PAT mode GRAPPA Accel. factor PE 4 Ref. lines PE 50 Reference scan mode Integrated Image Filter Off Distortion Corr. On TD 0.0 ms Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter Off Geometry Nr. of slice groups 1 Slices 25 Dist. factor 25 % Position R20.5 P23.7 F142.2 mm Phase enc. dir. R >> L Phase oversampling 100 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. SPAIR Water suppr. None Special sat. None Fat sat. mode Weak Special sat. None Set-n-Go Protocol On Table position P Inline Composing Off Restore magn. Off Tim CT mode Off System Body Off BO1 On BO2 On BO3 Off BO1 Off BO2 Off BO3 Off SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Adaptive Combine AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R20.5 P23.7 F142.2 mm Rotation 87.00 deg A >> P 250 mm R >> L 250 mm F >> H 124 mm Frequency 1H 123.200000 MHz Correction factor 1 optfsOptfs 1H 135.123 V Gain High Table position 142 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Dark blood Off Trajectory Cartesian Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Dimension 2D Compensate T2 decay Off Averaging mode Long term Multi-slice mode Interleaved Reduce Motion Sens. On Contrasts 1 Bandwidth 300 Hz/Px Flow comp. No Allowed delay 30 s Echo spacing 9.68 ms Define Turbo factor Turbo factor 24 Echo trains per slice 7 RF pulse type Low SAR Gradient mode Normal Hyperecho Off TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;SP4,5 Acquisition duration 0 ms Mode Min flip angle Min flip angle 150 deg BOLD Subtract Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off Contrasts 1 Save original images On Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\With Contrast\ep2d_dti_12 direction_p2 TA:4:11 PAT:2 Voxel size:2.5×2.5×5.0 mm Rel. SNR:1.00 :epse --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display Off Wait for user to start Off Start measurements single Routine Nr. of slice groups 1 Slices 30 Dist. factor 0 % Position R15.2 P29.6 F0.5 mm Orientation Transversal Phase enc. dir. A >> P AutoAlign --- Phase oversampling 0 % FoV read 320 mm FoV phase 81.5 % Slice thickness 5.0 mm TR 4700 ms TE 69.0 ms Averages 2 Concatenations 1 Filter Raw filter Coil elements BO1,2;BO3;SP4,5 Contrast MTC Off Magn. preparation None Fat suppr. SPAIR Fat sat. mode Strong Averaging mode Long term Delay in TR 0 ms Reconstruction Magnitude Multiple series Off Resolution Base resolution 130 Phase resolution 100 % Phase partial Fourier Off Interpolation Off PAT mode GRAPPA Accel. factor PE 2 Ref. lines PE 24 Reference scan mode Separate Distortion Corr. Off Prescan Normalize Off Normalize Off Raw filter On Intensity Weak Slope 25 Elliptical filter Off Dynamic Field Corr. Off Geometry Nr. of slice groups 1 Slices 30 Dist. factor 0 % Position R15.2 P29.6 F117.5 mm Phase enc. dir. A >> P Phase oversampling 0 % Multi-slice mode Interleaved Series Interleaved Nr. of sat. regions 0 Position mode L-P-H Fat suppr. SPAIR Special sat. None Fat sat. mode Strong Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off BO1 On BO2 On BO3 Off BO1 Off BO2 Off BO3 On SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 Off SP3 Off SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal R >> L Coronal A >> P Transversal F >> H Coil Combine Mode Sum of Squares AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment Off Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V ! Position R15.2 P29.6 F117.5 mm ! Rotation -3.20 deg ! R >> L 320 mm ! A >> P 197 mm ! F >> H 150 mm Frequency 1H 123.200000 MHz Correction factor 1 ExtExciteRF 1H 140.463 V Gain High Table position 117 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Magn. preparation None Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction On Averaging mode Long term Multi-slice mode Interleaved Bandwidth 2136 Hz/Px Optimization None Free echo spacing Off Echo spacing 0.59 ms EPI factor 106 RF pulse type Normal Gradient mode Fast* TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;BO3;SP4,5 Acquisition duration 0 ms BOLD Delay in TR 0 ms Diffusion mode MDDW Diff. weightings 3 b-value 1 0 s/mm² Diff. weighted images On Trace weighted images On ADC maps On FA maps On Mosaic On Tensor On Distortion Corr. Off b-Value >= 0 s/mm² Exponential ADC Maps Off Invert Gray Scale Off Calculated Image Off Calculated bValue 1400 s/mm² SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\With Contrast\T1 vibe fs Pre TA:2:25 PAT:8 Voxel size:1.0×1.0×1.5 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display On Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position R17.0 P21.2 H0.1 mm Orientation C > S4.1 Phase enc. dir. R >> L AutoAlign --- Phase oversampling 50 % Slice oversampling 0.0 % FoV read 360 mm FoV phase 100.0 % Slice thickness 1.50 mm TR 5.01 ms TE 1.73 ms Averages 5 Concatenations 1 Filter Distortion Corr.(2D), Prescan Normalize, Image Filter Coil elements BO1,2;BO2,3;SP2-5 Contrast Flip angle 10.0 deg Fat suppr. Water excit. fast Water suppr. None Dixon No Dixon Save original images On Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 352 Phase resolution 100 % Phase partial Fourier Off Interpolation Off PAT mode GRAPPA Accel. factor PE 4 Ref. lines PE 24 Reference scan mode Integrated Image Filter On Intensity Medium Edge Enhancement 3 Smoothing 3 Unfiltered images On Distortion Corr. On Accel. factor 3D 2 Ref. lines 3D 24 Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter Off POCS Read & Phase Slice resolution 50 % Slice partial Fourier Off Geometry Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position R17.0 P21.2 F50.9 mm Phase enc. dir. R >> L Phase oversampling 50 % Slice oversampling 0.0 % Slices per slab 160 Multi-slice mode Sequential Series Ascending Nr. of sat. regions 0 Position mode L-P-H Fat suppr. Water excit. fast Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off BO1 On BO2 On BO3 Off BO1 Off BO2 On BO3 On SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 On SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Sum of Squares AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment On Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R17.0 P21.2 F50.9 mm Rotation 0.00 deg F >> H 360 mm R >> L 360 mm A >> P 240 mm Frequency 1H 123.200000 MHz Correction factor 1 ExcitWEs 0 1H 91.768 V Gain Low Table position 51 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction Off Dimension 3D Elliptical scanning Off Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Weak Contrasts 1 Bandwidth 590 Hz/Px Optimization Min. TE Allowed delay 30 s RF pulse type Fast Gradient mode Fast Excitation Slab-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;BO2,3;SP2-5 Acquisition duration 0 ms Mode Off BOLD 3D centric reordering Off Time to center 14.9 s Subtract Off Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off SIEMENS MAGNETOM Skyra syngo MR D11 --------------------------------------------------------------------------------- \\USER\Neurography_JHH_Skyra\Sacral Plexus\With Contrast\T1 vibe fs Post TA:2:25 PAT:8 Voxel size:1.0×1.0×1.5 mm Rel. SNR:1.00 :fl --------------------------------------------------------------------------------- Properties Prio Recon Off Before measurement After measurement Load to viewer On Inline movie Off Auto store images On Load to stamp segments On Load images to graphic segments Off Auto open inline display On Wait for user to start Off Start measurements single Routine Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position R17.0 P21.2 H0.1 mm Orientation C > S4.1 Phase enc. dir. R >> L AutoAlign --- Phase oversampling 50 % Slice oversampling 0.0 % FoV read 360 mm FoV phase 100.0 % Slice thickness 1.50 mm TR 5.01 ms TE 1.73 ms Averages 5 Concatenations 1 Filter Distortion Corr.(2D), Prescan Normalize, Image Filter Coil elements BO1,2;BO2,3;SP2-5 Contrast Flip angle 10.0 deg Fat suppr. Water excit. fast Water suppr. None Dixon No Dixon Save original images On Averaging mode Short term Measurements 1 Reconstruction Magnitude Multiple series Each measurement Resolution Base resolution 352 Phase resolution 100 % Phase partial Fourier Off Interpolation Off PAT mode GRAPPA Accel. factor PE 4 Ref. lines PE 24 Reference scan mode Integrated Image Filter On Intensity Medium Edge Enhancement 3 Smoothing 3 Unfiltered images On Distortion Corr. On Accel. factor 3D 2 Ref. lines 3D 24 Mode 2D Unfiltered images Off Unfiltered images Off Prescan Normalize On Normalize Off B1 filter Off Raw filter Off Elliptical filter Off POCS Read & Phase Slice resolution 50 % Slice partial Fourier Off Geometry Nr. of slab groups 1 Slabs 1 Dist. factor 20 % Position R17.0 P21.2 F50.9 mm Phase enc. dir. R >> L Phase oversampling 50 % Slice oversampling 0.0 % Slices per slab 160 Multi-slice mode Sequential Series Ascending Nr. of sat. regions 0 Position mode L-P-H Fat suppr. Water excit. fast Water suppr. None Special sat. None Special sat. None Set-n-Go Protocol Off Table position P Inline Composing Off System Body Off BO1 On BO2 On BO3 Off BO1 Off BO2 On BO3 On SP5 On SP6 Off SP7 Off SP8 Off SP1 Off SP2 On SP3 On SP4 On Position mode L-P-H Positioning mode ISO Table position F Table position 0 mm MSMA S - C - T Sagittal L >> R Coronal P >> A Transversal H >> F Save uncombined Off Coil Combine Mode Sum of Squares AutoAlign --- Auto Coil Select Default Shim mode Standard Adjust with body coil Off Confirm freq. adjustment On Assume Dominant Fat Off Assume Silicone Off Adjustment Tolerance Auto ? Ref. amplitude 1H 0.000 V Position R17.0 P21.2 F50.9 mm Rotation 0.00 deg F >> H 360 mm R >> L 360 mm A >> P 240 mm Frequency 1H 123.200000 MHz Correction factor 1 ExcitWEs 0 1H 91.768 V Gain Low Table position 51 mm Img. Scale. Cor. 1.000 Physio 1st Signal/Mode None Resp. control Off Inline Inline Composing Off Distortion correction Off Sequence Introduction Off Dimension 3D Elliptical scanning Off Averaging mode Short term Multi-slice mode Sequential Asymmetric echo Weak Contrasts 1 Bandwidth 590 Hz/Px Optimization Min. TE Allowed delay 30 s RF pulse type Fast Gradient mode Fast Excitation Slab-sel. RF spoiling On TX/RX delta frequency 0 Hz TX Nucleus None TX delta frequency 0 Hz Coil elements BO1,2;BO2,3;SP2-5 Acquisition duration 0 ms Mode Off BOLD 3D centric reordering Off Time to center 14.9 s Subtract Off Liver registration Off Save images On Autoscaling Off Scaling factor 1 Offset 0 Subtrahend 1 Subtraction indices StdDev Off Std-Dev-Sag Off Std-Dev-Cor Off Std-Dev-Tra Off Std-Dev-Time Off MIP-Sag Off MIP-Cor Off MIP-Tra Off MIP-Time Off Radial MIP Off Save original images On Distortion Corr. On Mode 2D Unfiltered images Off MapIt None Contrasts 1 Save original images On Wash - In Off Wash - Out Off TTP Off PEI Off MIP - time Off Number of radial views 1 Axis of radial views L-R MPR Sag Off MPR Cor Off MPR Tra Off